6
•
Operation
Page 116
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Emptying the Seal Wash System
1. Unscrew the cap of the seal wash reservoir.
2. Together with the cap, remove the seal wash line from the seal wash
reservoir.
3. Remove the peristaltic tubing from the seal wash pump. When the
tubing is removed, any liquid that may be present in the seal wash
line, flows off to waste.
6.10.3 Restart after Long-Term Shutdown
To restart the device after a long-term shutdown, follow these steps:
1. Engage the peristaltic tubing in the seal wash pump.
2. Reconnect the seal wash line and the solvent lines to the
appropriate reservoirs.
3. Retighten the reservoir cap hand-tight. Verify that the retaining
guide remains in the hole in the reservoir cap. If it does not, press
the retaining guide into the hole to ensure that the tubing is kept in
place in the cap.
4. Turn on the device. Observe the following:
Situation
Action
If the device remained in the system
stack and all system modules were
turned off.
Turn on the system with the system
power button on the system base.
If the device is restarted after transport.
Turn on the device with the main power
switch.
5. Prepare and restart the other modules in the Vanquish system,
following the instructions in the
Operating Manuals
for the modules.
Pay special attention to the
Preparing the Module for Operation
section.
6. Purge the pump and then have the pump deliver for a short time.
7. Before starting an analysis, let the device equilibrate and be sure
that it is ready for operation. See
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...