6
•
Operation
Page 110
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
6.9
Optimizing the Performance of the Device
This section provides information for best performance of the device
and gives hints on what you can do to optimize the performance further.
6.9.1
General Guidelines
To optimize the performance of the device, consider the following
general guidelines:
•
Monitor the usage of specific device components that are subject to
wear and stress and schedule appropriate maintenance intervals
(see
).
•
Observe the general guidelines and recommendations on the use of
solvents and additives in the chromatography system (see
•
Pumps with built-in vacuum degasser:
Always use the built-in
vacuum degasser.
•
Purge the pump when changing to a different type of solvent or
when restarting the pump after periods of inactivity.
TIP
Shut off valves for solvent lines are available as an option (see
Installing Solvent Shut Off Valves (
). These valves allow you
to shut off the solvent flow through the system at the solvent reservoir,
for example, to prevent solvent from flowing through the system when
the pump flow is zero for a longer period.
6.9.2
Installing Solvent Shut Off Valves
When
To shut off the solvent flow through the system at the solvent reservoir,
for example, to prevent solvent from flowing through the system when
you open a flow connection on the low-pressure side
The steps in this section apply only when installing a shut off valve
after
initial operation of the pump. To install a shut off valve
during
initial
setup of the flow connections, follow the steps in
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...