Approximate Drying Times

Solvent

Volume/

Number

Time to dryness at drying rate:

Tubes

of tubes

LOW

MEDIUM

HIGH

Ethanol

0.5 ml

12

60± 5 min. 35± 5 min. 25± 5 min.

Table 1:

Ethanol drying in the Thermo Scientific Savant DNA SpeedVac. Drying times

presented as a guide only. Each system may vary slightly. The solvents used, and type
and quantity of solutes, also effect drying time. Thermo Scientific Savant DNA SpeedVac
Test Conditions: RD24 rotor with 12 x 1.5 ml tubes.

2.0

INSTALLATION (cont’d)

SET-UP

• Install the system on a stable surface that is clean, dry, level, and
within 4 feet (1.2 meters) of a compatible electrical outlet.

• Connect power cord to receptacle on the left side of the instrument

and plug into appropriate outlet. The cover latch mechanism releases
automatically.

WARNING: Before connecting the DNA120 to an outlet, make certain that

voltage,frequency,

and

amperage

match

the

DNA120

requirements indicated on the label on the left-hand side of the
instrument (120VAC / 60 Hz, 10 A or 230 VAC / 50 Hz, 5 A).

If there are any questions, please consult an electrician.

• Place rotor on the drive shaft in the sample chamber.

Hand-tighten

the hold-down knob until firmly seated.

CAUTION: Use only Thermo Scientific designed rotors in the

Savant DNA SpeedVac

®

.

For 230 VAC / 50 Hz, 5 A units only:

A safety latch prevents

the Savant DNA SpeedVac lid from operating until power is applied to
the system.

Do not attempt to open the lid unless the unit is

plugged in.

3.0

OPERATION

The switches on the front panel control operation of the Thermo Scientific Savant
DNA SpeedVac. The LED readout displays “run” parameters in minutes. You may
choose between three drying rates (LOW/MEDIUM/HIGH).

Summary of Contents for Thermo Scientific Savant SpeedVac DNA120

Page 1: ...Instruction Manual Thermo Scientific Savant DNA120 DNA120 OP SpeedVac Concentrators ...

Page 2: ... 2008 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Page 3: ...Section Title Operation Page 3 0 Operation 2 3 3 1 Mode Selection 4 3 2 Setting Target Time 4 3 3 Example of Manual Run 5 3 4 Example of Auto Run 6 3 5 Drying Rate 6 3 6 Other Topics 7 Section Section Title Specifications Page 4 0 Specifications 8 Section Section Title Accessories Page 5 0 Accessories 9 Appendix 1 Proper handling of hazardous exhaust vapors 10 Appendix 2 Operating hints 10 ...

Page 4: ...hanol precipitates of DNA RNA isopropanol water mixtures amplified reaction products in buffer aliquots of radiochemicals or other low volume samples see section 5 A unique swing out rotor RD2MP for microplates is available as well The Thermo Scientific Savant DNA SpeedVac is equipped with polypropylene vacuum fittings to avoid leaks and resist corrosion A chemical trap kit DTK120R and disposable ...

Page 5: ...latch mechanism releases automatically WARNING Before connecting the DNA120 to an outlet make certain that voltage frequency and amperage match the DNA120 requirements indicated on the label on the left hand side of the instrument 120VAC 60 Hz 10 A or 230 VAC 50 Hz 5 A If there are any questions please consult an electrician Place rotor on the drive shaft in the sample chamber Hand tighten the hol...

Page 6: ...d USING THE CONTROL PANEL A OPERATING MODE SWITCH see section 3 1 AUTO Press to start a run with automatic shut off OFF Press to stop any run MANUAL Press to start a run with manual shut off B TIMER SELECTION SWITCH see section 3 1 HEATER Press to select Heater timer RUN Press to select Run timer C CHANGE SETTING SWITCH see section 3 2 Press to increase value of timer selected by switch B Press to...

Page 7: ...operation of the Savant DNA120 in one of the two modes Select switch A to either MANUAL to initiate a concentrator run that you can start and stop manually The run starts when switch A is set to MANUAL when switch A is set to OFF The unit will end the concentrator run or AUTO to start a concentrator run for a predefined time interval The run starts when switch A is set to AUTO The run stops automa...

Page 8: ...r time than specified run time is specified then the HEATER shuts off at the end of the run Start the run by setting switch A to MANUAL During the run you can toggle the unit to display the following information Elapsed run time Set switch B to RUN Since there is no present run time the display counts upward showing the hours and minutes that have elapsed in the current run or Remaining heating ti...

Page 9: ...tes the amount of time remaining before the HEATER shuts off To terminate a run set switch A to OFF 3 4 EXAMPLE OF AN AUTO RUN Thermal energy can be applied to the concentrator chamber to counteract the cooling effect of evaporation it maintains samples in the liquid state and accelerates the concentration run The DRYING RATE switch selects the desired drying rate LOW maintains the concentrator ch...

Page 10: ...24 rotor 36 x 1 5 ml tubes RD36 rotor or 48 x 0 5 ml tubes RD48 rotor 72 x 0 5 ml tubes RD72 rotor or 2 microplates RD2MP rotor Chamber temperature Controlled at ambient 43 C or 65 C with over temperature safety shut off Weight 86 lbs 39 kg Dimensions W x D x H 11 in x 25 in x 11 in 29 cm x 63 cm x 29 cm Pump oil free diaphragm Displacement 36 l min at 60 Hz 30 l min at 50 Hz Ultimate Vacuum Level...

Page 11: ...izing solution 4 bottles for DNA120 OP GSC100 Glass safety cover standard on DNA120 OP DTK120R Chemical Trap Kit for processing radioactive samples connects to DNA SpeedVac exhaust port Includes SCT120 Chemical Trap tubing DC120R Disposable Cartridge DC120R Disposable Cartridge for trapping volatile radioactivity DC120R 4 Disposable Cartridge for trapping radioactivity package of four 5 0 ACCESSOR...

Page 12: ...uestions regarding a specific application contact Thermo Scientific APPENDIX 1 PROPER HANDLING OF HAZARDOUS EXHAUST VAPORS 1 Prewarm the chamber set switch B Heater to med setting 10 to 15 minutes prior to placing samples into the rotor 2 When the total volume of sample exceeds 6 to 8 ml per run a small amount of aerosol is created at exhaust port Use the post trap assembly to collect the condensa...

Page 13: ...er www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fischer Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europei Questo prodotto verrà smaltito o riciclato tramite ...

Page 14: ...ve information can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 1 866 984 3766 866 9 THERMO LABORATORY PARTS and SERVICE Phone 1 800 438 4851 TECHNICAL SUPPORT Phone 1 800 438 4851 ...

Page 15: ......

Page 16: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 104 3003 00 Rev D ...

Reviews: