5
Routine and Preventive Maintenance
Maintaining the ESI Probe
58
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
5. Turn the locking plate clockwise to the open position, and then carefully remove the ESI
probe from the probe heater. See
.
Figure 36.
Removing the ESI probe from the probe heater
Removing the ESI Capillary
To clean or replace the ESI capillary or to replace the ceramic sleeve, you must remove the ESI
capillary from the probe. See
for the part numbers of the ESI probe
components.
Y
To remove the capillary from the ESI probe
1. Remove the ESI probe from the MS detector as described in
.
CAUTION
Because its capillary tip and ceramic sleeve are fragile, exercise care when
you remove the ESI probe from the probe heater.
Locking plate
(in open position)
Probe heater
Source housing