background image

23

Unpacking and Installation

Warning

Use a properly grounded electrical outlet 
of correct voltage and current handling 
capacity.

Do not remove or modify grounded power 
plug. Use only properly grounded outlets 

to avoid shock hazard. Not rated for use in 
hazardous atmospheres.

Do not use in the presence of flammable or 
combustible materials; fire or explosion may 

result. This device contains components 
which may ignite such materials. If 
explosion proof models are required, 
contact Customer Service at 1-800-553-
0039 for more information.

Do not use in highly corrosive atmos-
pheres; corrosive fumes and spills may 
damage top and internal components, 

creating shock hazard.

Note

Some misalignment of the motor bearings 

in this product may have occurred during 
shipping. Prior to using this product, run the 
stirrer at maximum speed for ten minutes to 

realign the bearings.

Unpacking

Remove your SUPER-NUOVA

+

 unit from the carton. 

Inspect to ensure that the unit has not been damaged 

during shipment. If the unit appears to have sustained 

shipping damage contact the distributor from whom you 

purchased this product or Customer Service at 800-553- 

0039. Check for remote probe and thumbscrew prior to 

discarding packaging.

The following items are included in the shipment: 

SUPER-NUOVA

+

 unit

Cord

Remote Probe – PT100

Stir Bar (Stirring models only)

Knob  

Operation's Manual 

If any of these items are missing from the carton, 
contact Customer Service.

Installation

Set the unit on a flat stable surface at least 12" away 
from combustible materials, and plug the cord set into a 

properly grounded electrical outlet of correct voltage and 
current handling capacity.

Summary of Contents for SUPER-NUOVA+ HP88850190

Page 1: ...1 SUPER NUOVA Hot Plates Stirrers and Stirring Hotplates OPERATION MANUAL AND PARTS LIST Thermo S C I E N T I F I C CIC0000810 03 09 15...

Page 2: ...240V 7x7 Ceramic Stirring Hotplate HP88857194 100 120V 7x7 Aluminum Hot Plate HP88857198 220 240V 7x7 Aluminum Hot Plate SP88857194 100 120V 7x7 Aluminum Stirring Hotplate SP88857198 220 240V 7x7 Alum...

Page 3: ...t staff can supply technical information about proper setup operation or troubleshooting of your equipment We can fill your needs for spare or replacement parts or provide you with on site service We...

Page 4: ...ramic Hot Plate Specifications 18 Ceramic Stirrer Specifications 18 Ceramic Stirring Hotplate Specifications 19 Aluminum Hot Plate Specifications 19 Heating Specifications 20 Stirring Speed Specificat...

Page 5: ...etal Vessels and Sand Baths 30 Setting the Time 30 Hardware Over Temperature Protection 30 System Configuration 31 Temp Unit 32 Over temperature Set Point OTP 32 Preset Continue 33 Temperature Probe C...

Page 6: ...fication and remain at that temperature indefinitely Under these conditions the material being heated on the surface of the hotplate could reach levels in excess of the maximum temperature This manual...

Page 7: ...hazard If cord appears damaged replace immediately If cord is repeatedly loosened it is recommended to purchase the power entry module part number CIC0000793 or CIC0000469 depend on models listed in...

Page 8: ...superficie superior podr a elevarse a una temperatura m xima de 300 540 C dependiendo de las especificaciones de su modelo y permanecer a esa temperatura indefinidamente En estas condiciones el materi...

Page 9: ...ial de incendio Sustituya el cable inmediatamente si el cable est da ado Si el cable se afloja continuamente se recomienda comprar un m dulo de entrada de alimentaci n n mero de pieza CIC0000793 o dep...

Page 10: ...rfl chentemperatur auf die Maximaltemperatur ansteigen 300 540 C je nach Modellspezifikation und f r unbestimmte Zeit auf dieser Temperatur verbleiben Unter diesen Bedingungen kann das zu erhitzende M...

Page 11: ...ken schlagen und eine potentielle Brandgefahr darstellen Wenn das Kabel besch digt erscheint sofort ersetzen Wenn das Kabel sich wiederholt lockert wird empfohlen das Netzeingangsmodul Teilenummer CIC...

Page 12: ...ugmenter jusqu la temp rature maximale 300 540 C en fonction des sp cifications de votre mod le et demeurer ind finiment cette temp rature Dans ces conditions le mat riel chauff sur la surface de la p...

Page 13: ...intes reprises il est conseill d acheter le module d entr e de l alimentation num ro de pi ce CIC0000793 ou CIC0000469 selon les mod les inventori dans la section Pi ces de Rechange de ce manuel 7 Ne...

Page 14: ...superior pode aumentar para a temperatura m xima 300 540 C dependendo das especifica es do seu modelo e manter se na mesma durante um tempo indefinido Sob estas condi es o material a ser aquecido na...

Page 15: ...e e ou faiscar sendo um perigo potencial de inc ndio Se o cabo estiver danificado substitua de imediato Se o cabo estiver solto v rias vezes recomendamos que compre um adaptador de energia n mero de p...

Page 16: ...16 300 540 C 1 2...

Page 17: ...17 1 2 3 4 5 6 CIC0000793 CIC0000469 7 8 9 35 15 9 10 10 25 11 3 7 x 7 10 50 C 11 12 13 14...

Page 18: ...100 120 220 240 100 120 220 240 Amps 12 3 4 4 13 2 6 0 Watts 1480 1056 1580 1440 Freq 50 60 50 60 50 60 50 60 Phase 1 1 1 1 Max Temp C F 450 C 842 F 450 C 842 F 400 C 752 F 400 C 752 F Overall Dimensi...

Page 19: ...0 Amps 12 3 4 4 13 2 6 0 Watts 1480 1056 1580 1440 Freq 50 60 50 60 50 60 50 60 Phase 1 1 1 1 Max Temp C F 450 C 842 F 450 C 842 F 400 C 752 F 400 C 752 F Overall Dimensions HP88857194 HP88857198 SP88...

Page 20: ...C unloaded Temperature stability at the center of the top plate surface 100 C unloaded Accuracy of remote probe at user selected calibration temperature after calibration procedure Temperature stabili...

Page 21: ...e Part I General Requirements EN 61010 2 010 Part II Particular requirements for laboratory equipment for the heating of materials EN 61010 2 051 Part II Particular requirements for laboratory equipme...

Page 22: ...m and are suitable for use with glass or metal vessels and sand baths Other features of the SUPER NUOVA include a time feature that can be used to shut off heating Specific temperature setpoints may b...

Page 23: ...is product may have occurred during shipping Prior to using this product run the stirrer at maximum speed for ten minutes to realign the bearings Unpacking Remove your SUPER NUOVA unit from the carton...

Page 24: ...24 UNPACKING AND INSTALLATION SUPER NUOVA Hot Plate Control Panel SUPER NUOVA Stirrer Control Panel SUPER NUOVA Stirring Hot Plate Control Panel...

Page 25: ...25 UNPACKING AND INSTALLATION Control Panel Key Description 1 CAUTION HOT TOP Indicator 2 Display 3 Knob 4 Temperature Set Button 5 Speed Set Button 6 Time Set Button 7 Program Buttons...

Page 26: ...mperature OTP and the external probe Both top plate surface and OTP temperature are measured with independent analog circuits This provides a significant increase in safety due to the redundancy of th...

Page 27: ...the speed setpoint will no longer flash indicating that it has been set Once the unit begins stirring the display will indicate both actual speed and setpoint To stop the stirring action without chan...

Page 28: ...l accept any ungrounded PT100 PT1000 probe however the accuracy of probes other than those tested and specified may not reach the probe accuracy rating designated To set the temperature press TEMP key...

Page 29: ...top plate and the maximum top plate temperature 300 C 400 C or 450 C may be reached If the remote probe does not sense a temperature change in three minutes the unit will display ExProbe Err and shut...

Page 30: ...s the TIME key or wait 10 seconds the time setpoint will no longer flash indicating that it has been set In the temperature rising mode when the temperature is more than 2 degree below the temperature...

Page 31: ...ch type of unit the feature is applicable HP hot plate SP stirring hot plate S stir plate and the last column gives a brief description of the purpose of the feature Feature Display Availability Brief...

Page 32: ...perature Set Point OTP This option is only available for units that have heating functionality The Over temperature Set Point is the set point of the independent temperature control The purpose of the...

Page 33: ...guration by pressing and holding the knob After about 5 seconds a single you may remove your finger from the knob The display indicate system config on the first row the Temp Unit will show on the sec...

Page 34: ...f the display Turn the knob clockwise until the Calibrate Temp show on the second row of the display If you do no action the display will return to the operation status The option Calibrate Temp will...

Page 35: ...attached to element Ensure proper connection and polarity Check Heater Excessive top heat up time Internal thermocouple short circuit Failure in Internal thermocouple Failure in Element Failure in opt...

Page 36: ...36 7x7 Hot Plate Exploded View Exploded Views...

Page 37: ...37 EXPLODED VIEWS 7x7 Hot Plate Top Assembly...

Page 38: ...38 EXPLODED VIEWS 10x10 Hot Plate Exploded...

Page 39: ...39 EXPLODED VIEWS 10x10 Hot Plate Top Assembly...

Page 40: ...CONTROL PCB CIC0000573 CIC0000582 PC1 HP CONTROL PCB CIC0000574 CIC0000583 PC1 S CONTROL PCB CIC0000589 CIC0000590 PC2 DISPLAY PCB CIC0000575 CIC0000584 S1 SPEED SENSOR SC1313X1 SC1313X1 S2 ENCODER SW...

Page 41: ...r 11 SC1313X1 SC1313X1 Speed Sensor 12 CIC0000482 CIC0000484 Motor 120V 12 CIC0000483 CIC0000485 Motor 230V 13 JNX35 Lower Insulation 14 JNX36 JNX38 Upper Insulation 15 CIC0000541 CIC0000543 Heating E...

Page 42: ...Probe PT1000 Uncoated CIC0000588 Exterior Probe Hastelloy PT1000 7077 Thermometer Clamp 7068 90 Clamp Holder 1000 2 12 Aluminum Rod 7078 Large Clamp 7079 Small Clamp up to 1 2 CIC0001146 Transparent...

Page 43: ...t beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned...