Maintenance and Care
5. 6. Autoclaving
Always disassemble all parts before autoclaving, e.g. lids need to be removed before autoclaving a bucket or
rotor.
If not stated otherwise on the parts themselves, all parts can be autoclaved at 121 °C for 20 min. Refer to
for details on rotors.
Make sure that the necessary sterility is achieved according to your requirements.
After autoclaving, treat the entire surface of aluminum parts including the cavities with corrosion protection oil
(70009824).
Treat the bolts of swing out rotors with bolt grease (75003786) if necessary.
CAUTION
Never exceed the permitted temperature and duration when autoclaving.
NOTICE
No chemical additives are permitted in the steam.
5. 7. Maintenance
Lifetime
The centrifuge is specified for a lifetime of 10 years. Decommissioning the centrifuge is suggested when this
limit is reached.
The lifetime of a rotor and buckets depends on the amount of physical load. Do not use rotors and buckets that
exceed the maximum number of cycles. The maximum number of cycles for rotors and buckets is given in the
chapter
“Rotor Specifications” on page B-1
. The maximum number of cycles for buckets is marked on the
buckets themselves. Fiberlite rotors are not limited in cycles, but have a limited lifetime of 15 years.
Preventive Maintenance
In order to keep this product able to perform the intended applications reliably and safely, ongoing preventive
maintenance is necessary in accordance with the following recommended schedule:
The anti-vibration mounts (20038955) and the motor cover (20058551) need to be replaced every 3 years.
The gas spring (50154683) and the particle deflection seal (50154683) need to be checked annually and
replaced if necessary. Refer to
For rotors and buckets mind the information in
“Rotor and Accessories Inspection” on page 5-1
CAUTION
Usage beyond these limits might affect the safety of the overall system.
NOTICE
In the worst case the centrifuge, the used accessories and the samples
can be damaged.
NOTICE
Maintenance activities must be performed by Thermo Fisher Scientific
authorized service technicians only.
Summary of Contents for Sorvall ST Plus Series
Page 116: ...B 9 TX 750 B 1 4 Biocontainment Certificate ...
Page 120: ...B 13 TX 1000 B 2 4 Biocontainment Certificate ...
Page 126: ...B 19 HIGHConic II B 4 4 Biocontainment Certificate ...
Page 133: ...B 26 Fiberlite F15 6 x 100y B 7 4 Biocontainment Certificate ...
Page 137: ...B 30 HIGHPlate 6000 B 9 4 Biocontainment Certificate ...
Page 140: ...B 33 M 20 Microplate B 10 4 Biocontainment Certificate ...
Page 143: ...B 36 Microliter 48 x 2 B 11 4 Biocontainment Certificate ...
Page 146: ...B 39 Microliter 30 x 2 B 12 4 Biocontainment Certificate ...
Page 149: ...B 42 MicroClick 30 x 2 B 13 4 Biocontainment Certificate ...
Page 152: ...B 45 MicroClick 18 x 5 B 14 4 Biocontainment Certificate ...
Page 155: ...B 48 Fiberlite F21 48 x 2 B 15 4 Biocontainment Certificate ...
Page 159: ...B 52 Fiberlite F10 6 x 100 LEX B 17 4 Biocontainment Certificate ...