3-10
Graphical User Interface
3. 2. 4. Set Temperature
A refrigerated centrifuge allows for preselecting a temperature for the sample between -10 °C and +40 °C for
the centrifugation run. This feature is not available on ventilated models.
A
CAUTION
Due to air friction the temperature of the rotor may rise significantly while the centrifuge is
spinning. Displayed and set temperature can deviate from sample temperature. Sample temperature might
exceed critical temperature of your apllication.
Proceed as follows to set a temperature for the next centrifugation run:
1. Tap the
temperature box
on the “Home” screen, shown in
Figure 3–21: Temperature Box on Home Screen
The “Setpoints: Standard” screen or “Setpoints: Advanced” screen is displayed.
2. Tap in the
Temp
entry field, then use the keypad on the right to enter the temperature for the sample setpoint.
The admissible temperature range for the unit appears just below the “Temp” label to assist you in setting
the temperature setpoint correctly.
When you tap the
Temperature
entry field, the previous setting will be replaced immediately as you start
entering digits on the keypad.
3. Tap the
Save
button to save the setpoint temperature for the next centrifugation run.
Summary of Contents for Sorvall ST Plus Series
Page 116: ...B 9 TX 750 B 1 4 Biocontainment Certificate ...
Page 120: ...B 13 TX 1000 B 2 4 Biocontainment Certificate ...
Page 126: ...B 19 HIGHConic II B 4 4 Biocontainment Certificate ...
Page 133: ...B 26 Fiberlite F15 6 x 100y B 7 4 Biocontainment Certificate ...
Page 137: ...B 30 HIGHPlate 6000 B 9 4 Biocontainment Certificate ...
Page 140: ...B 33 M 20 Microplate B 10 4 Biocontainment Certificate ...
Page 143: ...B 36 Microliter 48 x 2 B 11 4 Biocontainment Certificate ...
Page 146: ...B 39 Microliter 30 x 2 B 12 4 Biocontainment Certificate ...
Page 149: ...B 42 MicroClick 30 x 2 B 13 4 Biocontainment Certificate ...
Page 152: ...B 45 MicroClick 18 x 5 B 14 4 Biocontainment Certificate ...
Page 155: ...B 48 Fiberlite F21 48 x 2 B 15 4 Biocontainment Certificate ...
Page 159: ...B 52 Fiberlite F10 6 x 100 LEX B 17 4 Biocontainment Certificate ...