4
Operation
Saving programs
4-6
ST16 / 16R
Thermo Scientific
Prewarming or precooling the centrifuge
For setting the pretemp value for the centrifuge proceed as follows:
1. Press the key
in order to open the temperature selection menu.
The display shows the message "Set PreTemp".
2. Enter the desired value by pressing the key
repeatedly, until the desired value shows.
3. The display shows:
The display shows the current temperature inside the rotor chamber.
4. Press the START key
.
The rotor chamber is cooled down or heated up to the preset temperature.
5. Press the STOP key
.
The display shows the current temperature inside the rotor chamber.
Bucket selection
Bucket selection is only possible for swing-out rotors. The bucket code corresponds to the last four
digits of the bucket catalog number.
1. Press the BUCKET key
.
The display shows the following message:
2. Press the BUCKET key
repeatedly until the bucket being used is displayed.
Saving programs
1. Enter the program parameters.
2. Press any of the program store keys for 4 seconds.