background image

Thermo Scientific

 

Diagnostik Rotor

Instruction

 Manual

50135351-

b

  •  0

7

 / 2020

Visit us online to register your warranty:

thermofisher.com/labwarranty

Summary of Contents for Sorvall RC-4

Page 1: ...Thermo Scientific Diagnostik Rotor Instruction Manual 50135351 b 07 2020 Visit us online to register your warranty thermofisher com labwarranty ...

Page 2: ...WEEE Conformity This product is subject to the regulations of the EU Waste Electrical Electronic Equipment WEEE Directive 2012 19 EU It is marked by the following symbol ...

Page 3: ...ter 3 Rotor Installation 3 1 Rotor Installation 3 2 Removing the Rotor 3 2 Chapter 4 Rotor Loading 4 1 Before a Run 4 2 Proper Loading 4 2 Improper Loading 4 3 Chapter 5 Maintenance and Care 5 1 Cleaning Intervals 5 2 Cleaning 5 2 Disinfection 5 3 Decontamination 5 4 Service of Thermo Fisher Scientific 5 5 RCF Values A 1 Chemical Compatibility Chart B 1 Table of Contents ...

Page 4: ...C Contents ii 75006480 DiagnostikTM Rotor Thermo Scientific ...

Page 5: ... in this instruction manual will result in the expiration of the sellers warranty Scope of Supply If any parts are missing please contact your nearest Thermo Fisher Scientific representative Article Number Quantity Check 75006480 Diagnostik Rotor 1 75003786 Bolt Grease 1 50134706 CD with Manual 1 This symbol refers to general hazards WARNING means that injuries or material damage or contamination ...

Page 6: ...ad the rotor Operate the rotor always with the lid closed Use only accessories which have been approved by Thermo Fisher Scientific Exceptions to this rule are commercially available glass or plastic centrifuge tubes provided they have been approved for the speed or the RCF value of the rotor Please observe the safety instructions Please pay particular attention to the following aspects Rotor inst...

Page 7: ...Thermo Scientific 75006480 DiagnostikTM Rotor 1 1 1 Rotor Data Contents Technical Data on page 1 2 ...

Page 8: ...Acceleration Braking Time s 65 65 65 65 60 65 Sample Heating at nmax C referred to Ambient Temperature of 23 25 C Run Time 60 Minutes 0 0 0 Max Speed nmax rpm at 4 C 3500 3500 3500 Aerosol tight No No No Autoclavable No No No Table 1 2 200 V 50 Hz Centrifuge Cryofuge 5500i Catalog Number 75004479 Weight of empty Rotor kg 10 7 Maximum permissible Load g 2 x 2000 Maximum Speed nmax rpm 3500 Maximum ...

Page 9: ...ture of 23 25 C Run Time 60 Minutes 0 Max Speed nmax rpm at 4 C 3500 Aerosol tight No Autoclavable No Table 1 4 230 V 50 60 Hz Centrifuge Sorvall RC 4 Sorvall RC 4 Catalog Number 75004473 75004481 Weight of empty Rotor kg 10 7 10 7 Maximum permissible Load g 2 x 2000 2 x 2000 Maximum Speed nmax rpm 3500 3500 Maximum RCF Value at nmax 2547 2547 Radius max min cm 18 6 14 1 18 6 14 1 Acceleration Bra...

Page 10: ...ting at nmax C referred to Ambient Temperature of 23 25 C Run Time 60 Minutes 0 Max Speed nmax rpm at 4 C 3500 Aerosol tight No Autoclavable No Table 1 6 120 V 60 Hz Centrifuge RC 4 Catalog Number 75004477 Weight of empty Rotor kg 10 7 Maximum permissible Load g 2 x 2000 Maximum Speed nmax rpm 3500 Maximum RCF Value at nmax 2547 Radius max min cm 18 6 14 1 Acceleration Braking Time s 65 65 Sample ...

Page 11: ...Thermo Scientific 75006480 DiagnostikTM Rotor 2 1 2 Accessories Contents Accessory on page 2 2 ...

Page 12: ...ccessories for Rotor 75006480 Adapter for Diagnostik Rack Rotor Length Width mm Maximum Height incl Tubes and Rack mm Racks per Rotor Article Number Hitachi 118 x 20 120 2 x 10 75006416 Olympus 177 x 20 120 2 x 6 75006417 Sysmex 200 x 25 120 2 x 6 75006418 Dade Behring 120 2 x 2 75006419 ...

Page 13: ...Thermo Scientific 75006480 DiagnostikTM Rotor 3 1 3 Rotor Installation Contents Rotor Installation on page 3 2 Removing the Rotor on page 3 2 ...

Page 14: ...f available close the rotor with the rotor lid Removing the Rotor To remove the rotor proceed as follows 1 Open the centrifuge door 2 Thread the fastening tool in the centrifuge spindle counter clockwise Hold the rotor with the other hand into position 3 At the same time pull the rotor directly upwards with both hands and remove it from the centrifuge spindle Make sure not to tilt the rotor while ...

Page 15: ...Thermo Scientific 75006480 DiagnostikTM Rotor 4 1 4 Rotor Loading Contents Before a Run on page 4 2 Proper Loading on page 4 2 Improper Loading on page 4 3 ...

Page 16: ... bottles do not touch the bucket lid 6 Check the rotor bolts and apply grease 75003786 before using for the first time 7 Check that each bucket or micro plate carrier can swing freely by moving it carefully with your hand Weigh the bucket content adapter and tube Make sure you do not exceed the maximum compartment load Proper Loading To ensure safe operation of the centrifuge the rotor must be eve...

Page 17: ...4 Rotor Loading Improper Loading Thermo Scientific 75006480 DiagnostikTM Rotor 4 3 Improper Loading ...

Page 18: ...4 Rotor Loading Improper Loading 4 4 75006480 DiagnostikTM Rotor Thermo Scientific ...

Page 19: ...gnostikTM Rotor 5 1 5 Maintenance and Care Contents Cleaning Intervals on page 5 2 Cleaning on page 5 2 Disinfection on page 5 3 Decontamination on page 5 4 Autoklavieren auf Seite 6 5 Service of Thermo Fisher Scientific on page 5 5 ...

Page 20: ...emperature must never exceed 50 C since higher temperatures could damage the material and shorten the lifespan of the parts Use only disinfectants with a pH of 6 8 Dry aluminum parts off with a soft cloth After cleaning treat the entire surface of aluminum parts with corrosion protection oil 70009824 Also treat the cavities with oil Store the aluminum parts at room temperature or in a cold storage...

Page 21: ...ice in the rotor chamber Let the ice melt and drain it off Clean the rotor chamber as described above Disinfection Disinfect the centrifuge immediately whenever infectious material has spilled during centrifugation CAUTION Before using any cleaning or decontamination methods except those recommended by the manufacturer users should check with the manufacturer that the proposed method will not dama...

Page 22: ... 9 Dispose of the disinfectant according to the applicable guidelines 10 Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a maximum temperature of 50 C 11 After cleaning treat the entire surface of aluminum parts with corrosion protection oil 70009824 Also treat the cavities with oil 12 Treat the bolt of the swing out rotor with bolt grease 75003786 Dec...

Page 23: ...um temperature of 50 C 11 After cleaning treat the entire surface of aluminum parts with corrosion protection oil 70009824 Also treat the cavities with oil 12 Treat the bolt of the swing out rotor with bolt grease 75003786 Service of Thermo Fisher Scientific Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an authorized service technician The servic...

Page 24: ...5 Maintenance and Care Service of Thermo Fisher Scientific 5 6 75006480 DiagnostikTM Rotor Thermo Scientific ...

Page 25: ...100 18 6 14 1 252 191 1200 18 6 14 1 299 227 1300 18 6 14 1 351 266 1400 18 6 14 1 408 309 1500 18 6 14 1 468 355 1600 18 6 14 1 532 404 1700 18 6 14 1 601 456 1800 18 6 14 1 674 511 1900 18 6 14 1 751 569 2000 18 6 14 1 832 631 2100 18 6 14 1 917 695 2200 18 6 14 1 1006 763 2300 18 6 14 1 1100 834 2400 18 6 14 1 1198 908 2500 18 6 14 1 1300 985 2600 18 6 14 1 1406 1066 2700 18 6 14 1 1516 1149 28...

Page 26: ... A 2 75006480 DiagnostikTM Rotor Thermo Scientific 3100 18 6 14 1 1998 1515 3200 18 6 14 1 2129 1614 3300 18 6 14 1 2265 1717 3400 18 6 14 1 2404 1822 3500 18 6 14 1 2547 1931 Speed rpm Rmin Rmax RCF Rmin RCF Rmax ...

Page 27: ...U U S S Formic Acid 100 S M U U U S M U U S S U S U S U Ammonium Acetate S S U S S S S S S S S S S U S S S S S S S S S S Ammonium Carbonate M S U S S S S S S S S S S S U U S S S S S S M S S S Ammonium Hydroxide 10 U U S U S S M S S S S S S U M S S S S S S S S S M S Ammonium Hydroxide 28 U U S U S U M S S S S S U S U M S S S S S S S S S M S Ammonium Hydroxide conc U U U U S U M S S S U S U U S S S ...

Page 28: ...M S U U S S M M S U U S U M U U M M M U S S S S U U Ferric Chloride U U S M S M S S S S M U S S Acetic Acid Glacial S S U U S S U M S U S U U U U U M S U M U S U U S U Acetic Acid 5 S S M S S S M S S S S S M S S S S S S S M S S M S S M Acetic Acid 60 S S U U S S U S M S U U M U S M S M S M S M U S M U Ethyl Acetate M M U U S S M M S S U S U M U U S S U U S M M S U U Ethyl Alcohol 50 S S S S S S M ...

Page 29: ...M M S M S S M S S S S M S M U S S Kerosene S S S S S S U S M U S U M M S M M M S S U S S U S Sodium Chloride 10 S S S S S S S S S S S S S S S S S S M S Sodium Chloride sat d U S U S S S S S S S S S S S S S M S Carbon Tetrachloride U U M S S U M U S U U S U M U S S M M S M M M M U S S Aqua Regia U U U U U U U U U U U S M Solution 555 20 S S S S S S S S S S S S S S S S S S S S Magnesium Chloride M S...

Page 30: ...S Perchloric Acid 10 U U S U U S M M M U M S M M M S U S S Perchloric Acid 70 U U U U U S U M U U M U U U M M U M S U U S U S Phenol 5 U S U S M M S U M U U S U M S M S U U S U M M M S Phenol 50 U S U S U M S U M U U U U U S U M U U S U U U M S Phosphoric Acid 10 U U M S S S U S S S S U S S S S S S S S S U M U S S Phosphoric Acid conc U U M M U S M S U U M M S S S M S M S U M U S Physiologic Media...

Page 31: ... U U U S U U U U U U U U U S U U S Trisodium Phosphate S M S S S S S S S Tris Buffer neutral pH U S S S S S S S S S S S S S S S S S S S S S S S S S Triton X 100 S S S S S S S S S S S S S S S S S S S S S S S S S Urea S U S S S S S S S M S S S S S S S M S S Hydrogen Peroxide 10 U U M S S U U S S S U S S S M U S S S S S S M S U S Hydrogen Peroxide 3 S M S S S S S S S S S S S S M S S S S S S S S S S X...

Page 32: ...s for materials under the stress of centrifugation When in doubt we recommend pretesting sample lots Key S Satisfactory M Moderate attack may be satisfactory for use in centrifuge depending on length of exposure speed involved etc Suggest testing under actual conditions of use U Unsatisfactory not recommended Performance unknown suggest testing using sample to avoid loss of valuable material ...

Page 33: ......

Page 34: ... consult your local sales represen tative for details Shown pictures within the manual are examples and may differ considering the set parameters and language Pictures of the user interface within the manual are showing the English version as example Italy 39 02 95059 552 Japan 81 3 5826 1616 Korea 82 2 2023 0600 Netherlands 31 76 579 55 55 New Zealand 64 9 980 6700 Nordic Baltic CIS countries 358...

Reviews: