37 / 94
Technical Specifications
17R Series Centrifuges – PCR 4 x 8 Rotor
Voltage
230 V
120 V
Maximum Speed
13 300
13 300
Maximum RCF-Value
13 100
13 100
Acceleration / Braking Time
9 s / 12 s
9 s / 12 s
Min. Temperature at max. Speed, Ambient
Temperature of 23 °C
≤ 0 °C
≤ 0 °C
21R Series Centrifuges – PCR 4 x 8 Rotor
Voltage
230 V
120 V
Maximum Speed
14 800
14 800
Maximum RCF-Value
16 200
16 200
Acceleration / Braking Time
11 s / 13 s
11 s / 13 s
Min. Temperature at max. Speed, Ambient
Temperature of 23 °C
≤ 0 °C
≤ 0 °C
Table 32:
PCR 4 x 8 Rotor Performance Data
Accessories
Description
Article
No�
Rotor
Capacity
Max� Tube
Dimensions
ClickSeal biocontainment lid
75003410
-
-
Replacement O-ring set for ClickSeal lid –
75003410
75003405
-
-
Table 33:
Accessories PCR 4 x 8 Rotor
1. 6. 7. Hematocrit Rotor
For more information on the hematocrit rotor refer to the separate rotor manual.