![Thermo Scientific SOLA II Flare User Manual Download Page 39](http://html1.mh-extra.com/html/thermo-scientific/sola-ii-flare/sola-ii-flare_user-manual_1105273039.webp)
Chapter 3
Start-Up & Shutdown
Thermo Scientific
SOLA II Flare User Manual
3-3
7.
Enable analysis of the sample by taking the instrument out of the
suspend mode. The unit begins injecting and analyzing sample when
the instrument temperatures reach operating levels.
8.
Allow the analyzer system to stabilize. Monitor the measured values on
the front panel for consistent analysis readings to determine when the
analyzer system has stabilized.
Startup After Short-
Term Shutdown
If initially starting the analyzer or if major service was performed, use the initial
start-up procedure detailed in the previous section,
When starting the analyzer after a short-term shutdown, perform the following:
1.
Open the instrument and purge air and carrier air to the instrument.
Refer to the applications information, manuals, and drawings shipped
with the instrument, and adjust the pressure regulators to the
appropriate settings.
2.
Apply power to the analyzer.
Note
If the X-Purge option is installed, you must follow the instructions
▲
3.
Allow the analyzer to warm up until the instrument stabilizes.
4.
Adjust the flow rates.
5.
Turn on sample to the instrument.
The analyzer includes an interlock system that prevents the injection valve from
operating until the pyrolyzer and oven temperatures reach the operating value.
Short-Term
Shutdown
Follow this procedure when temporarily shutting down the analyzer. To shut
down the analyzer for maintenance purposes or for a long-term shutdown, refer
to the following section,
1.
Observe reported sulfur value.
2.
Do not interrupt power, instrument air, carrier or auxiliary air, or open
oven doors until reported sulfur value is less than 0.5 ppm or reported
sulfur value has not changed by more than 2% for 15 minutes.
Note
If possible, switch sample selection to nitrogen to sweep the system.
▲
Summary of Contents for SOLA II Flare
Page 1: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0...
Page 2: ......
Page 3: ...SOLA II Flare Sulfur Online Analyzer User Manual PN 90 1312 0 Revision A...
Page 4: ......
Page 6: ......
Page 8: ......
Page 14: ......
Page 18: ......
Page 44: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 80: ......
Page 104: ......
Page 122: ......
Page 136: ......
Page 148: ......
Page 158: ......
Page 160: ......
Page 162: ......