Thermo Scientific Revco Slimline Value Freezers

Installation and Operation

6

9 Chart Recorders 

Panel-mounted six-inch seven-day recorders are available as 
options for all freezer models.

9.1 Set Up and Operation

To prepare the recorder to function properly, complete the 

following steps:

1. Open the recorder door to access the recorder.
2. Connect the nine volt DC battery located at the recorder’s 

upper right corner. This battery provides backup power.

3. Install clean chart paper (refer to Section 9.3 below).
4. Remove the plastic cap from the pen stylus and close the 

recorder door.

Recorder operation begins when the system is powered on.The 
recorder may not respond until the system reaches temperatures 
within the recorder’s range. 

Figure 3.  Chart Recorder

Figure 4.  Chart Buttons

9 Volt Battery

Imprinting
Stylus

Hub-Nut and
Retaining Wire

Chart

Chart Buttons

Reference 
Mark

1

2

3

CHANGE CHART

9.2 Power Supply

The recorder normally uses AC power when the system is 
operating. If AC power fails, the LED indicator flashes to alert 
you to a power failure. The recorder continues sensing cabinet 
temperature and the chart continues turning for approximately 

24 hours with back-up power provided by the nine-volt battery. 

The LED indicator glows continuously when main power is 
functioning and the battery is charged.

When the battery is low, the LED flashes to indicate that the 
battery needs to be changed.

9.3 Changing Chart Paper

To change the chart paper, complete the following steps:

1. Locate the pressure sensitive buttons at the front, upper left 

of the recorder panel.

2. Press and hold the Change Chart button (#3) for one second. 

The pen will move off the scale.

3. Unscrew the center nut, remove the old chart paper, and 

install new chart paper. Carefully align the day and time with 

the reference mark (a small groove on the left side of the 
recorder panel).

4. Replace the center nut and hand tighten. Press the Change 

Chart button again to resume temperature recording.

9.4 Calibration Adjustment

This recorder has been accurately calibrated at the factory and 
retains calibration even during power interruptions. If required, 

however, adjustments can be made as follows:

1. Run the unit continuously at the control setpoint temperature. 

Continue steady operation for at least two hours to provide 

adequate time for recorder response.

2. Measure cabinet center temperature with a calibrated 

temperature monitor.

3. Compare the recorder temperature to the measured cabinet 

temperature. If necessary, adjust recorder by pressing the left 
(#1) and right (#2) chart buttons.

Note:

The stylus does not begin to move until the button is held 

for five seconds.

Summary of Contents for Revco Slimline

Page 1: ...Installation and Operation Manual Thermo Scientific Revco Slimline Value Ultra Low Temperature SI Freezers ...

Page 2: ... registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries 2007 Thermo Fisher Scientific All rights reserved Suva is a registered trademark of DuPont All other trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries ...

Page 3: ...Operation 4 7 1 Control Panel Features 4 7 2 Start Up 4 7 2 1 Turning the Power On 4 7 2 2 Setting the Cabinet Temperature 4 7 2 3 Alarm 4 8 Maintenance and Troubleshooting 5 8 1 Condenser Maintenance 5 8 1 1 Cleaning the Condenser 5 8 1 2 Cleaning the Condenser Filter 5 8 2 Gasket Maintenance 5 8 3 Defrost Procedures 5 8 4 Alarm Battery Maintenance 5 9 Chart Recorders 6 9 1 Set Up and Operation 6...

Page 4: ...y Your unit must be properly grounded in conformity with national and local electrical codes Never connect the unit to overloaded power sources Disconnect the unit from all power sources before cleaning troubleshooting or performing other maintenance on the product or its controls 3 Pre Installation 3 1 Unpacking At delivery examine the exterior for physical damage while the carrier s representati...

Page 5: ...afety and trouble free operation this unit must be properly grounded before it is used Failure to ground the equipment may cause personal injury or damage to the equipment Always conform to the National Electrical Code and local codes Do not connect the unit to overloaded power lines Your freezer is equipped with a service cord and plug designed to connect it to a power outlet which delivers the c...

Page 6: ...within 30 seconds after door openings To minimize warm air entry the port automatically closes as soon as pressures are equalized The pressure equalization port is located on the right side of the storage chamber Although the port is designed to self defrost excessive frost accumulation in the bottom of the chamber could eventually restrict air flow Therefore you should periodically inspect the po...

Page 7: ... 1 Insert a small screwdriver into the slotted screw labeled Control Setpoint and simultaneously press and hold the Setpoint Display button The temperature display changes to read the existing setpoint value 2 Turn the setpoint screw clockwise for a colder setting and counterclockwise for a warmer setting until the desired setpoint shows in the digital temperature display 3 Release the Setpoint Di...

Page 8: ... 4 on page 2 for leveling information Keep the lid and door gaskets clean and frost free by wiping gently with a soft cloth 8 3 Defrosting the Freezer Defrost the freezer once or year or whenever the ice buildup exceeds 3 8 To defrost complete the following steps 1 Remove all products and place in another cabinet 2 Turn off the freezer 3 Open the outer door and all inner doors 4 Let the freezer st...

Page 9: ...ery The LED indicator glows continuously when main power is functioning and the battery is charged When the battery is low the LED flashes to indicate that the battery needs to be changed 9 3 Changing Chart Paper To change the chart paper complete the following steps 1 Locate the pressure sensitive buttons at the front upper left of the recorder panel 2 Press and hold the Change Chart button 3 for...

Page 10: ... is required Chart Paper Surge Suppressor Alarm Systems Remote Alarm Monitoring Description Catalog No For All Models Fiberboard boxes and grid dividers are packaged in lots of one dozen Fiberboard box 5 1 4 in square x 2 in high one dozen 5954 Fiberboard box 5 1 4 in square x 3 in high one dozen 5956 Grid divider fiberboard 100 cell 7 16 in holds 100 12 mm vials one dozen 5958 Grid divider fiberb...

Page 11: ...en die unter diese RoHS Anweisungfallen finden Sie unter www thermo com Conformità WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE È marcato col seguente simbolo Thermo Fisher Scientific ha stipulato contratti con una o diverse società di riciclaggio smaltimento in ognuno degli Stati Membri Europe...

Page 12: ... can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number IF YOU NEED ASSISTANCE SALES DIVISION Phone 828 658 2711 800 252 7100 FAX 828 645 3368 LABORATORY PARTS and SERVICE Phone 800 438 4851 FAX 828 658 2576 TECHNICAL SUPPORT Phone 800 438 4851 ...

Page 13: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 310655H01 Rev C ...

Reviews: