Startup & Shutdown
Procedure
7-10
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
Thermo Fisher Scientific
The Heaters tab (
) allows for adjustment and setting of
the source temperature, inlet probe heater power, RMS body
temperature, and RMS sample tube temperature.
Figure 7
–11.
Heaters tab
Open the Instrument Status window (see
). The tabs and
the readings displayed under each tab are listed below.
Analyser tab
Cabinet temperature: Within the range 10°C to 35°C.
Electronics temperature: Within the range 10°C to 45°C.
System pressure: Typically 5 x 10
-6
mbar, but application
dependent. Refer to installation records for setting of
optimum pressure.
Source tab
Ion energy: Approximately 1000 V (for 150 amu mass range)
or 800 V (for 200 amu mass range).
Electron energy: Typically set to 70 V.
Filament current: Depends on filament type:
2.4-2.9 A for a Thoria coated iridium filament
1.5-2.5 A for a W/ThO coiled filament
3.0- 3.8 A for a Re filament
Note
If filament is close to zero, check that Trap Current is
set to at least 20 µA and EE is at least 50 V.
Heaters
Checking
Instrument
Status
Parameters