![Thermo Scientific Orion Lab Star DO113 Operation Manual Download Page 35](http://html.mh-extra.com/html/thermo-scientific/orion-lab-star-do113/orion-lab-star-do113_operation-manual_1105222035.webp)
Orion Lab Star DO113 Dissolved Oxygen Meter 35
Chapter 7
Meter Info
7.1 Meter Specifications
Dissolved Oxygen as Percent Saturation
Range
0.0 to 300.0 % saturation
Resolution
0.1, 1 % saturation
Relative accuracy ±2 % saturation ± 1 LSD
Calibration types
100 % saturation and 0 % saturation
Barometric
pressure
compensation
400.0 to 850.0 mmHg; 0.533 to
1.133 bar
Salinity
compensation
0.0 to 50.0 ppt
Dissolved Oxygen as Concentration
Units
mg/L or ppm
Range
0.00 to 30.0 mg/L or ppm
Resolution
0.01, 0.1 mg/L or ppm
Relative accuracy ±0.2 mg/L or ppm ± 1 LSD
Calibration points 1 point, custom concentration value
Barometric
pressure
compensation
400.0 to 850.0 mmHg; 0.533 to
1.133 bar
Salinity
compensation
0.0 to 50.0 ppt
Temperature
Range
-5.0 to 105.0 ºC, 23.0 to 221.0 ºF
Resolution
0.1 ºC, 0.1 ºF
Relative accuracy ±0.3 ºC, ±0.5 ºF
Calibration points 1 point
Offset adjustment Up to ±5 ºC, up to ±9 ºF
Features
Display
5” backlit LCD
Time and Date
Yes
Time format
12 hour AM/PM or 24 hour
Date format
Selectable, DD-MM-YYYY or MM-
DD-YYYY
Read types
Continuous, Auto-Read, Timed
Timed interval
5 seconds to 60 minutes
Data log
500 data sets with time and date
Calibration log
Active DO as % saturation, DO as
concentration, temperature with time
and date
Calibration alarm
Yes, 1 to 168 hours or off
Sleep mode
Yes, optional
Self-test
Automatic with startup
Memory
Non-volatile
Input
9-pin MiniDIN
Data output
Computer or printer
Output format
CSV or print
Warranty
3 years
Certifications
CE, TUV 3-1, FCC Class A
Enclosure
IP-54
Power
100-240 VAC, 50-60Hz, 9 DC
adapter, 1.3A
Dimensions
(L x W x H)
187.2 mm x 155.2 mm x 61.7 mm,
7.37” x 6.11” x 2.43”
Weight
700 grams, 1.54 lbs.
Environmental conditions
Environmental
conditions
Indoors
Altitude
Up to 2,000 Meters
Operating
temperature
5°C to 45°C
Operating relative
humidity
5 to 85%, non-condensing
Storage
temperature
-20°C to 60°C
Storage relative
humidity
5 to 85%, non-condensing
Mains fluctuation ±10% of range (100-240VAC)
Installation
category
II
Pollution degree
2
Protection class
III
Summary of Contents for Orion Lab Star DO113