Concentration Measurements—Standard Curve Application
Measuring Samples
94
Orion AquaMate 7000 Vis and Orion AquaMate 8000 UV-Vis
Thermo Scientific
If a 6-Position Cell Holder is installed, press the cell position buttons to reposition the cell
holder and measure the rest of the standards manually.
When all of the standards have been measured, the Standards screen (see
) shows the absorbance of each standard, along with the
slope, intercept and correlation coefficient of the standard curve.
Using the Standards Screen
Here is an example of the Standards screen showing measurement results:
You can use this screen for:
• Displaying a graph of the standard curve data (press
View Graph
)
•
(press
Save Test
)
•
(press
Edit Standards
)
•
(press
Run Test
)
Measuring Samples
To measure samples automatically using the calibration curve (using Auto 6 or
Auto 3)
1. Press
Run Test
.
2. Install the blank and samples.
3. Press
Enter
.
The Standard Curve screen shows the absorbance and concentration of each sample.
To switch between tabular and graphical displays, press
View Graph/Tabular
.