6
External Devices
Communication Connectors
18
Orbitrap Tribrid Series Getting Connected Guide
Thermo Scientific
Communication Connectors
Use the left-side communication connectors to connect the MS to the Ethernet switch, the
syringe pump, the modular valve, and any external device.
lists the pin-out descriptions for the connectors shown in
.
IMPORTANT
• Before proceeding, verify that the external device is suitable for use with the
Orbitrap Tribrid Series MS. Make sure that the signal from the external device meets
the requirement stated in the pin 4 (
pin-out description (see
). If it
does not meet this requirement, you cannot use it with the MS.
• To help prevent sample waste if the MS becomes unresponsive, connect a cable
between the optional LC device and the MS’s Ready Out and Injection Hold pins (see
). Then, configure the Contact Closure settings in the Method Editor. You are
responsible for providing this cable.
CAUTION
After connecting the external device cables, attach the connector housings.
Table 3.
Pin-out descriptions for the communication connectors (Sheet 1 of 3)
Pin
Name
Description
–
Reset
Resets the instrument to a power-up state.
Note
Use this button only if the instrument does not
respond to the control program on the data system
computer or if you need to restart the instrument
without turning off the electronics service switch.
Peripheral Control
1
Ground
Earth ground.
2
5V
Provides a 5 Vdc, 500 mA output (with pin 1).