12
Error Codes
below lists the error messages that may appear in the control panel display window
(see
) and provides instructions for clearing such alarms.
Table 12-1 Heratherm Oven Error Codes
Error Message & Code
Root Cause
Alarm Response
Alarm Clearing Instructions
1
Display Error (E002)
Display communication
error. The built-in control-
ler was unable to restore
communication with the
control panel.
Audible alarm activated,
alarm relay energized,
message shown on dis-
play. Reset after 30 s.
Power cycle the device by
unplugging, then recon-
necting the power cord. If
this doesn’t solve the
problem, call service.
Mirrored Parameter
Loaded (E003)
The controller was unable
to read the user-specific
settings and had to resort
to an emergency parame-
ter set held in mirrored
storage.
Audible alarm activated,
alarm relay energized.
Fallback to mirrored
parameter storage.
Device continues to run
without loss of functional-
ity, including user-specific
settings.
Check the latest settings,
for example the set value.
Factory Parameter
Loaded (E004)
The controller was unable
to read the mirrored
parameter set and had to
resort to factory-preset
parameters.
Fallback to factory-preset
parameters. Audible alarm
activated, alarm relay
energized, message
shown on display. User-
specific settings may be
lost - for example, the
temperature display unit
preference, or user pro-
grams.
Acknowledge by pressing
. Re-enter customer-
specific settings
.
Default Parameter
Loaded (E005)
The controller was unable
to read the factory-preset
parameters and had to
resort to default settings
Fallback to default param-
eters. Audible alarm acti-
vated, alarm relay
energized, message
shown on display. The
device is completely inop-
erative.
Call service.
Power Down Error
(E007)
Power has been cut off
(power outage) while the
device was running.
Audible alarm activated,
alarm relay energized,
message shown on dis-
play.
Check the power sup-
ply.
Power up then
device, then acknowledge
the alarm by pressing
.
Summary of Contents for OGS 100
Page 1: ......
Page 6: ...Contents iv Heratherm Thermo Scientific...
Page 46: ...Product Description Tube Access Ports 4 16 Heratherm Thermo Scientific...
Page 60: ...Start up Connecting the RS 232 Interface 5 14 Heratherm Thermo Scientific...
Page 76: ...Handling and Control Switching the Oven Off Powering Down 7 14 Heratherm Thermo Scientific...
Page 78: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific...
Page 82: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific...
Page 88: ...Maintenance Returns for Repair 10 6 Heratherm Thermo Scientific...
Page 104: ...Spare Parts and Accessories 14 6 Heratherm Thermo Scientific...
Page 106: ...Device Log 15 2 Heratherm Thermo Scientific...
Page 108: ...Contact 16 2 Heratherm Thermo Scientific...