 Gently agitate aspirator 
bottle. Loosen clamp 
and cell suspension 
will flow into the Cell 
Factory. Chambers will 
initially fill unevenly 
due to a difference in 
pressure.

9

10

 When filling is 
complete, allow 
levels of liquid in 
all chambers to 
equalize through the 
connecting channel.

The surface of all the 
trays will be covered 
by medium. Remove 
the connector, replace 
the white filter adaptor 
cap, but leave the filter 
on and incubate.

13

14

To empty the Cell 
Factory place it above 
aspirator bottle, in the 
position shown, and 
the liquid will run out.

 Turn the Cell Factory 
90º, so that filling inlet 
is up. Medium will 
be separated with 
equal amounts in each 
chamber.

11

12

 Place the Cell Factory 
in horizontal incubation 
position.

Tilt the Cell Factory to 
drain it completely.

15

16

To decontaminate the 
Cell Factory after use 
it is recommended to 
autoclave: 60 min,  
132°C.

© 2010 Thermo Fisher Scientific Inc. All Rights Reserved.

All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

 MNLSPCF10INSTR 0210

47201 Ver. 6.0

Asia:

 China Toll-free: 800-810-5118 or 400-650-5118; India: +91 22 6716 2200, India Toll-free: 1 800 22 8374; Japan: +81 3 3816 3355

Other Asian countries: 65 68729717 

Europe:

 Austria: +43 1 801 40 0; Belgium: +32 53 73 42 41; Denmark: +45 4631 2000; France: +33 2 2803 2180; Germany: +49 6184 90 

6940, Germany Toll-free: 08001-536 376; Italy: +39 02 02 95059 or 434-254-375;  

Netherlands: +31 76 571 4440; Nordic/Baltic countries: +358 9 329 100; Russia/CIS: +7 (812) 703 42 15;  

Spain/Portugal: +34 93 223 09 18; Switzerland: +41 44 454 12 12; UK/Ireland: +44 870 609 9203 

North America:

 USA/1 585 586 8800; USA Toll-free: +1 800 625 4327

South America:

 USA sales support: +1 585 899 7198

Countries not listed:

 +49 6184 90 6940 or +33 2 2803 2180

www.thermoscientific.com/bioproduction

Thermo Scientific Nunc Cell Factory

Instructions for use: CF1 CF4, CF10 and CF40

Reviews: