Applications
Measure Custom
30
NanoDrop One
C
with NanoDropQC Software User Guide
Thermo Scientific
5. Select
Load Method
.
A message box shows the NanoDrop One methods available on the selected USB device.
6. Select one or more method names in the Load Method box to select the methods to load.
7. Select
Load
.
To measure using a custom method
Before you begin...
Before taking pedestal measurements with the NanoDrop One instrument, lift the instrument
arm and clean the upper and lower pedestals. At a minimum, wipe the pedestals with a new
laboratory wipe. For more information, see
.
To measure a sample using a custom method using the local instrument interface
1. Make sure the method resides in same location as the database where you want to store
the measurement results (see
2. From the Home screen, select
Custom Method
.
3. In the Select Method box, select to select the method to run.
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will flow into
the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will permanently
damage the quartz fiber optic cables.