Maintenance
Instrument Diagnostics
124
NanoDrop One
C
with NanoDropQC Software User Guide
Thermo Scientific
To interpret intensity check results
If one of these indicators:
• UV
• Visible
• Bias
has an adjacent yellow triangle instead of the green check marks shown above,
pedestals with deionized water
and then repeat the Intensity Check.
If a yellow triangle appears next to the Bias indicator, make sure the room is within the
temperature specifications for the instrument.
If the Intensity Check fails again,
.
Performance Verification
Run Performance Verification every 6 months to confirm pathlength accuracy is within
specifications.
Supplies needed
• lint-free laboratory wipes
• deionized water (DI H
2
O)
• calibrated precision pipettor (0–2 µL)
•
PV-1 performance verification solution
(liquid photometric standard available only from
us or a local distributor)
• laboratory gloves