background image

6

POWER REQUIREMENTS:

100, 100IDE, 105, 120, 125:

120VAC, 50/60 Hz, 100 Watts, 0.8 Amps

150:

120VAC, 50/60 Hz, 150 Watts, 1.3 Amps

150JPN:

100VAC, 50/60 Hz, 104 Watts, 1.0 Amps

101, 106, 121, 126:

240VAC, 50/60 Hz, 100 Watts, 0.4 Amps

152:

240VAC, 50/60 Hz, 150 Watts, 0.6 Amps

100JPN, 120JPN:

100VAC, 50/60 Hz,   70 Watts, 0.7 Amps

100-5JPN, 120-5JPN:

100VAC, 50/60 Hz, 100 Watts, 1.0 Amps

TEMPERATURE RANGE:

100, 100IDE, 100JPN, 101 105, 106:

From slightly above ambient to 40ºC

120, 120JPN, 125, 126:

From slightly above ambient to 60ºC

150, 150JPN, 152:

From slightly above ambient to 45ºC

100-5JPN, 120-5JPN:

From slightly above ambient to 65ºC

INTERIOR DIMENSIONS:

100(IDE), 100JPN, 100-5JPN, 101, 105, 106:

12"W x 12"D x 8"H

(30 x 30 x 20 cm)

120, 120JPN, 120-5JPN, 121, 125, 126:

12"W x 12"D x 12"H

(30 x 30 x 30 cm)

150, 150JPN, 152;

13-7/16"W x 13-11/32"D x 12-11/16"H

(36 x 34 x 32 cm)

EXTERIOR DIMENSIONS:

100(IDE), 100JPN, 100-5JPN, 101, 105, 106:

14-3/8"W x 13-11/32"D x 11-5/16"H

(37 x 34 x 30 cm)

120, 120JPN, 120-5JPN, 121, 125, 126:

14-3/8"W x 13-11/32"D x 15-5/16"H

(37 x 34 x 41 cm)

150, 150JPN, 152:

13-7/16"W x 13-11/32"D x 15-5/16"H

(37 x 34 x 41 cm)

CHAMBER VOLUME:

100(IDE), 100JPN, 100-5JPN, 101, 105, 106:

0.67 cubic feet (19 liters)

120, 120JPN, 120-5JPN, 121, 125, 126:

1.00 cubic feet (28 liters)

150, 150JPN, 152:

1.32 cubic feet (37 liters)

SHIPPING WEIGHT:

100(IDE), 100JPN, 100-5JPN, 101, 105, 106:

15 lbs. (7 kg)

120, 120JPN, 120-5JPN, 121,125, 126:

27 lbs. (12 kg)

150, 150JPN, 152:

33 lbs. (15 kg)

UNIT’S ENVIRONMENTAL OPERATING CONDITIONS:

POLLUTION DEGREE:

2

INSTALLATION CATEGORY:

II

ALTITUDE:

2000 Meters MSL (Mean Sea Level)

HUMIDITY:

80% maximum, non-condensing

ELECTRICAL SUPPLY:

120VAC or 240VAC

VOLTAGE TOLERANCE:

±10% of normal rated line

TEMPERATURE: 

15ºC to 40ºC

                           PRODUCT USAGE:

 This product is intended for use indoors 

only

SECTION 4

Summary of Contents for Lab-Line 100

Page 1: ...e Compact Low Cost Incubators OPERATION AND REPAIR MANUAL AND PARTS LIST Model Numbers 100 150 Lab Line 100JPN 150JPN 100IDE 152 100 5JPN 101 105 106 120 120JPN 120 5JPN 121 125 126 Additional comment...

Page 2: ...REV J LAB LINE COMPACT LOW COST INCUBATORS MODEL NO 100 100IDE 100JPN 100 5JPN 101 105 106 120 120JPN 120 5JPN 121 125 126 150 150JPN 152 1999 North 15th Ave Melrose Park IL 60160 1491 USA PHONE 563...

Page 3: ...3 TABLE OF CONTENTS SECTION TITLE 1 Introduction 2 Description 3 Specifications 4 Installation 5 Operation 6 Maintenance 7 Replacement Parts Warranty...

Page 4: ...his unit becomes familiar with the Operation Section of this manual Be sure to fill out the Warranty Registration Card and mail it in to Lab Line Instruments within seven 7 days after receiving the un...

Page 5: ...elp to distribute heat evenly throughout the chamber Powder coated durable heavy gauge steel exteriors make the units easy to clean and maintain A bimetallic thermostat controls temperature Models 105...

Page 6: ...JPN 152 13 7 16 W x 13 11 32 D x 12 11 16 H 36 x 34 x 32 cm EXTERIOR DIMENSIONS 100 IDE 100JPN 100 5JPN 101 105 106 14 3 8 W x 13 11 32 D x 11 5 16 H 37 x 34 x 30 cm 120 120JPN 120 5JPN 121 125 126 14...

Page 7: ...nit nameplate requirements Allow clearance around the unit for free air convection accessory attachment and user convenience The incubator door opens from the left side THERMOMETER INSTALLATION The th...

Page 8: ...Because of the variety of plug configurations in use worldwide for 240 VAC power the unit is furnished with the plug removed The user must install a plug to conform with local code and configuration r...

Page 9: ...ght hand side of incubator rotate the thermostat knob clockwise to increase the temperature counterclockwise to lower the temperature The status lamp is lit when the heaters are energized 2 Allow one...

Page 10: ...e plug from its outlet for a thorough cleaning between incubation periods The chamber interior and shelves can be cleaned with a mild detergent or soap and water Do not use abrasives or metallic scour...

Page 11: ...40V units 152 340 209 01 Tension Catch 600 124 00 Hinge 570 341 00 Rubber Feet 4 790 214 00 Shelf 100 105 120 125 Series 172 134 00 Shelf Lower 150 Series 172 135 00 Shelf Upper 120 125 150 Series 172...

Page 12: ...e products with the replacement parts but only for a period of ninety 90 days from the Commencement Date Lab Line s warranty provided hereunder shall be null and void and without further force or effe...

Page 13: ...ncubators Ovens Dual Action Open Air Shaker Reciprocating Shakers open air and water bath Rockers and Rotators Low Cost Shakers Environ Blok Shaker Titer Plate Shaker Multi Wrist Shaker Water Baths ex...

Page 14: ...15 FIRST IN INSTRUMENTS SERVICING SCIENCE INDUSTRY RESEARCH AND EDUCATION SINCE 1908...

Page 15: ...16 ACCESSORY CHECKLIST The following loose parts and accessories are packed with this unit Before discarding any packing materials please be sure that nothing has been overlooked...

Page 16: ...100 5JPN 101 105 106 120 120JPN 120 5JPN 121 125 126 172 134 00 _________ Shelf Lower 172 135 00 1 Models 150 150JPN 152 _________ Shelf Upper 172 081 00 1 Models 120 120JPN 120 5JPN 121 125 126 _____...

Reviews: