Installation
Installation setups
Thermo Fisher Scientific
Thermo Scientific KingFisher Flex User Manual 29
Figure 3
–18.
Shield plate fitted
Fastening/changing the heating block is controlled by the software. To fit
the heating block, choose the Change_heatblock protocol under the
Maintenance menu by using the up and down cursor keys. Then press
START
. The heating block will thus rise. Insert the heating block by first
setting it and then pressing it hard into place (
). The heating block will snap into place at both ends if you
have done it correctly.
Figure 3
–19.
Inserting the heating block
Heating block