2
Changing the Column
Replacing the Factory Installed Column
Thermo Scientific
ISQ User Guide
11
Figure 4.
Column Flowmeter Connector
b. Connect the flowmeter to the dedicated fitting on the column flowmeter connector.
c.
If you have a TRACE 1310, select the
Back Column
or
Front Column
button in
the
Configuration
menu. Otherwise, perform the column evaluation through the
Chromatography Data System. See the
TRACE 1300 and TRACE 1310 User Guide
for instructions.
d. Select
Column
and input the column’s physical characteristics.
e.
If a pre-/post column is present, set the length and nominal internal diameter of the
pre-/post column in the same valid ranges for the column.
f.
Start the column evaluation:
i.
On the column page, click
Column Evaluation
.
ii. Connect a flowmeter to the column flowmeter connector as indicated above.
iii. Press
Start
. The inlet is pressurized to deliver a 5 mL calculated column flow.
iv. Once the pressure has stabilized, measure the flow out of the column flowmeter
connector. Enter the actual flow when prompted. The corrected column ID will
be displayed.
v.
Select
Apply
and the K-factor will be adjusted to give the corrected column flow.
According to the physical characteristics of the column, the system calculates and
displays the relevant column K-factor. At the end of the routine, a message will
indicate that the evaluation was successful.
Note
For the most reproducible results, you should conduct a more detailed
column evaluation. However, the following steps, while recommended, are not
required.
A B C
To flowmeter