2
Changing the Column
Replacing the Factory Installed Column
10
ISQ User Guide
Thermo Scientific
d. Allow the column to purge for at least 10 minutes. If you used methanol to detect
column flow, remove column from methanol during purge time.
e.
Insert the column into the fitting of the column flowmeter connector that blocks the
column flow.
15. Perform a column leak check:
a. On the TRACE 1310, select the
Leak Check
icon in the
Maintenance
menu.
Otherwise, perform the leak check through the Chromatography Data System. Refer
to the
TRACE 1300 and TRACE 1310 Series GC User Guide
for instructions.
b. Start the leak check.
The split and purge valves of the selected channel are automatically closed, and the
channel is pressurized with carrier gas to the leak check setpoint.
The system monitors the pressure for one minute. If the pressure does not drop more
than the maximum allowed sensitivity value, then the leak check will pass. If the leak
check does not pass, use the leak detector to find and fix any leaks.
c.
Repeat the leak check until no leaks are indicated.
16. Calibrate the carrier gas flow (column evaluation):
a. Carefully push the capillary column end into the flowmeter section of the column
flowmeter connector. See
Note
If you are using a TRACE GC Ultra or a FOCUS GC, refer to your GC
user documentation for instructions on performing a leak check.
Tip
Leaks can be caused by not tightening the fitting on the column flowmeter
connector. Check the fitting before looking for the leak elsewhere.
CAUTION INSTRUMENT DAMAGE:
Do not allow the column flowmeter
connector to exceed 80 °C (176 °F). Otherwise, it will melt and damage the
instrument.
Summary of Contents for ISQ
Page 1: ...ISQ User Guide 1R120555 0003 Revision D January 2013 ...
Page 22: ......
Page 54: ......
Page 82: ...4 Creating a Method Creating a Method for the GC 60 ISQ User Guide Thermo Scientific ...
Page 124: ......
Page 148: ......
Page 152: ......