Consumables
Gas Supply
Thermo Scientific
iCAP RQ Pre-Installation Requirements Guide (P/N BRE0009927, Revision A)
5-7
Plasma and Cooling Gas
The argon supply to the iCAP RQ instrument must be stabilized. The
gas pressure regulator must be qualified for the purity specification of
the plasma gas, see
.
Using a high quality two-stage regulator is recommended, see
for suppliers of pressure regulators.
NOTICE
The plasma gas regulator must have an outlet pressure of max.
1 MPa (10 bar), and it must fit the maximum primary pressure
specification of the argon gas laboratory supply. Check whether your
maximum primary pressure is 30 MPa (300 bar), 23 MPa (230 bar), or
20 MPa (200 bar), and choose accordingly.
▲
Connection of the laboratory argon supply to the iCAP RQ instrument
is made via a push-in fitting. A recommended set-up for the argon
supply is shown in
. 3 m polyurethane tubing will be supplied
with the instrument.
Table 5-4.
Argon gas requirements
Gas Requirement
Specification
Purity
99.996% or better
Maximum Water Content
5 ppmv
Supply Rate
max. 24 L/min
Pressure
min. 0.55 MPa (5.5 bar)
max. 0.6 MPa (6 bar)
using a two-stage pressure
regulator (200 bar to 10 bar,
230 bar to 10 bar or 300 bar to
10 bar)