Maintenance
Maintaining the Inlet System
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
8-15
Cleaning the Injector
For the cleaning procedure of the spray chamber, see
Inserting the Injector
❖
To install the injector
1. Insert the injector and its holder in the torch assembly.
2. Insert the spray chamber, see
.
3. Insert the nebulizer, see
.
Maintaining the Torch
Removing the Torch
❖
To remove the injector
1. Remove the nebulizer, see
.
2. Remove the spray chamber, see
.
3. Carefully pull out the injector.
4. Press the red button at the torch holder (
3
in
the torch assembly counterclockwise to release the torch assembly.
5. Pull out the torch assembly.
6. Carefully remove the torch body from the torch holder.
Labeled Components: 1=torch body, 2=gas inlet, 3=release button, 4=torch
holder
Figure 8-6.
Torch disassembled
1
3
4
2
2