Installation
Room ventilation
3-2
Heratherm IMC 18
Thermo Scientific
Room ventilation
Heat dissipating from the device during continuous operation may cause a change in the room
climate.
Therefore, the device must only be installed in rooms with sufficient ventilation.
Do not install the incubator in room recesses without ventilation.
When several devices are to be placed in the same room, additional ventilation may have to be
provided as necessary.
The room ventilation is to be of a technical form which is designed in accordance with the thermal
input.
Installation
When installing the device, make sure that the installation and supply connections remain freely
accessible. The minimum wall clearance is 5 cm / 2 in on the sides and back. No other devices are to
be placed on the device.
Transport
The device is not to be lifted by the door or attachments when being transported.
The cooling fan on the back of the device is not to be covered.
It is not to suck-in any loose items (e.g. paper).
Summary of Contents for Heratherm IMC 18
Page 1: ......
Page 10: ...Safety Information Intended use of the device 1 6 Heratherm IMC 18 Thermo Scientific ...
Page 12: ...Delivery of the device Scope of supply 2 2 Heratherm IMC 18 Thermo Scientific ...
Page 18: ...Description of unit Condensation tray 4 4 Heratherm IMC 18 Thermo Scientific ...
Page 22: ...Operation Starting Operation 6 2 Heratherm IMC 18 Thermo Scientific ...
Page 26: ...Shut down Shutting the device down 8 2 Heratherm IMC 18 Thermo Scientific ...
Page 32: ...Maintenance Inspections and checks 10 2 Heratherm IMC 18 Thermo Scientific ...
Page 38: ...Contact Information 13 2 Heratherm IMC 18 Thermo Scientific ...
Page 39: ...Contact Information Thermo Scientific Heratherm IMC 18 13 3 ...
Page 40: ......