1
Introduction and Description
Directives, Standards and Guidelines
1-4
Heraeus Pico 17 / 21
Thermo Scientific
Directives, Standards and Guidelines
Height
Width
Depth
230 mm
240 mm
350 mm
230 mm
240 mm
350 mm
230 mm
240 mm
350 mm
Weight with empty Rotor
10.5 kg
10.5 kg
10.5 kg
Table 1-2.
Technical data
Heraeus Pico
21
Feature
Value
Table 1-3.
Directives, Standards and Guidelines
Region
Directive
Standard
Europe
220-230 V, 50 / 60 Hz
230 V, 50 / 60 Hz
98/79/EC In Vitro Diagnostics
2011/65/EU RoHS
Directive on the Restriction of the use of certain
Hazardous Substances in electrical and
electronic equipment
Protective go
al
s of:
2006/42/EC Machinery
2014/35/EU Low Voltage
2014/30/EU Electromagnetic Compatibility (EMC)
EN 61010-1
EN 61010-2-020
EN 61010-2-101
EN 61326-1 Class B
EN 61326-2-6
EN ISO 14971
EN ISO 13485
USA & Canada
220-230 V, 50 / 60 Hz
120 V, 60 Hz
FDA listed
Product code JQC
centrifuges for clinical use
Device class 1
ANSI/UL 61010-1
UL 61010-2-020
UL 61010-2-101
EN ISO 14971
EN ISO 13485
Japan
100 V, 50 / 60 Hz
China
230 V, 50 / 60 Hz
208-240 V, 50 / 60 Hz
CFDA listed
IEC 61010-1
IEC 61010-2-020
IEC 61010-2-101
IEC 61326-2-6
IEC 61326-1 Class B
EN ISO 14971
EN ISO 13485
Summary of Contents for Heraeus Pico 17
Page 30: ...4 10 Heraeus Pico 17 21 Thermo Scientific 4 Operation ...
Page 51: ......