![Thermo Scientific Heracell Vios 160i CR Operating Instructions Manual Download Page 81](http://html1.mh-extra.com/html/thermo-scientific/heracell-vios-160i-cr/heracell-vios-160i-cr_operating-instructions-manual_1104558081.webp)
Thermo Scientific
Heracell Vios 160i / 250i CR
5-1
5
Operation
Preparing the Device
The device must not be released for operation before all major start-up activities (see
) have been completed.
Device Check:
Prior to starting operation, the following device components must be checked for their correct function
• Gas hoses must be seated tight on the connecting filter and must be secured using a hose clamp.
• The access port must be capped.
• The pressure compensation opening must be permeable, its insert must be installed in the workspace.
• The glass door seal must not be damaged.
• The glass door/gas-tight screen measurement port must be capped.
• The shelf system components must be installed safely.
• The airbox with the HEPA filter and the air duct must be installed according to the instructions.
• Check the HEPA filter on the rear wall filter enclosure for correct fit. The red warning light on the
rear wall filter enclosure indicates a fault in the fan of the rear wall filter enclosure.
Sterilizing the Device Workspace:
• Run the Steri-Run sterilization routine (see
“Steri-Run Sterilization Routine”
disinfect the workspace according to the hygiene regulations set forth by the operator.
Table of Contents
•
•
CAUTION
Limited operating temperature for the Cell Locker Chambers!
Before a sterilization run the Cell Locker Chambers must be removed. The max. operating
temperature for the Cell Locker Chambers is 121°C/250 °F. The Cell Locker Chambers will melt if
they remain in the device during the Steri-Run sterilization cycle.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...