6
Handling and Control
User Configuration
6-26
Heracell Vios 160i / 250i CR
Thermo Scientific
Reminder Interval Setting
The reminder intervals are integral components of the alarm and monitoring system of the device
control. For the two essential functions Steri-Run and auto-start as well as for routine service work, the
user can set dates that trigger an alarm whenever they occur.
The counting begins at 00:00 hrs of the day on which the previously set reminder interval has elapsed.
On the due date, the display shows a reminder message for the activated reminder interval:
• Steri-Run: Please run Steri-Run routine.
• auto-start: Please run auto-start routine. Appears upon successful completion of the Steri-Run
sterilization routine.
• Service interval: Request service. The service message can be confirmed. In this case, the Request
Service icon is displayed.
After the routines have been run successfully, the reminder messages are hidden.
Factory Settings
Steri-Run sterilization routine
90 days
auto-start routine
Off
Service interval
Off
Inner chamber, HEPA filter interval
365 days
Reminder Interval Setting
1. Press the
Menu
key.
2. Select the Reminder intervals menu option.
• The selection dialog shown in
is displayed.
Figure 6-29.
Choose the Desired Reminder Interval Function.
3. Select the appropriate menu option, for example
Steri-Run
.
• The input dialog shown in
is displayed.
Summary of Contents for Heracell Vios 160i CR
Page 8: ...Table of Contents 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 12: ...Figures 4 Heracell Vios 160i 250i CR Thermo Scientific...
Page 86: ...5 Operation Water Filling 5 6 Heracell Vios 160i 250i CR Thermo Scientific...
Page 138: ...6 Handling and Control Error Messages 6 52 Heracell Vios 160i 250i CR Thermo Scientific...
Page 178: ...11 Specifications Heracell Vios 250i CR 11 10 Heracell Vios 160i 250i CR Thermo Scientific...
Page 208: ...13 Device Log 13 2 Heracell Vios 160i 250i CR Thermo Scientific...
Page 212: ...14 Contact Information Thermo Scientific 14 4 Heracell Vios 160i 250i CR Thermo Scientific...