12
Basic Concentration Measurements—Basic A-%T-C
Using Conc/Factor to Measure Concentration
Thermo Scientific
GENESYS 10S Vis User Guide
51
2. Press
Measure Blank
.
If the Single Cell Changer is installed, remove the blank and place the standard in the cell
changer.
3. Press
Units/Standard
to set the units and measure the standard.
4. Press
Enter Conc
, enter the concentration value of the standard and press
Enter
.
5. Press
Select Units
, highlight the appropriate unit in the list and press
Enter
.
6. Press
Measure Standard
.
The instrument measures the absorbance of the standard and displays the absorbance and
calculated factor.
Using Conc/Factor to Measure Concentration
In this mode a concentration factor is used to determine the concentration of samples. The
concentration of unknown samples is determined ratiometrically from the entered factor. The
measurement can be expressed mathematically as
(Abs / %T of Sample) * Factor = Concentration of Sample
where the factor is entered, the Abs/%T of the sample is measured, and the sample
concentration is calculated.
Y
To use Conc/Factor to measure concentration
1. If necessary, press
Change Mode
to switch to the Conc With Factor mode.
2. Press
Units/Factor
to set the factor and select the units.
3. Press
Enter Factor
.
4. Type the desired factor value.
Summary of Contents for GENESYS 10S Vis
Page 1: ...GENESYS 10S Vis User Guide 269 251700 Revision A October 2009 ...
Page 8: ...Preface viii GENESYS 10S Vis User Guide Thermo Scientific ...
Page 32: ...3 Accessories External Printers 24 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 44: ...6 Managing Stored Tests Deleting a Test 36 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 50: ...9 Calculator Function 42 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 82: ...16 Absorbance Ratio Measuring Samples 74 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 90: ...18 3 Point Net Taking Measurements 82 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 124: ...22 Maintenance Replacing the Tungsten Lamp 116 GENESYS 10S Vis User Guide Thermo Scientific ...
Page 134: ...23 Parameters 126 GENESYS 10S Vis User Guide Thermo Scientific ...