Problem Solving
Self-test Warning Messages
Thermo Scientific
G100 Range CO
2
Analyser
8-3
Change Battery
The Lithium Ion battery has exceeded its recommended
life-time or number of charge cycles and should be changed.
The instrument will need to be returned to the manufacturer
for service.
User Cal. Due
It has been over a month since the instrument was last user
calibrated. For optimal performance and accuracy it is rec-
ommended that the instrument is user calibrated each time it
is used.
Invalid Time
The instrument has an invalid time. This is most likely to
occur after a reset. The correct time should be entered using
the set ‘Time & Date’ option via the ‘Utilities’ menu.
Invalid Date
The instrument has an invalid date. This is most likely to
occur after a reset. The correct date should be entered using
the set ‘Time & Date’ option via the ‘Utilities’ menu.
Baro. Fault
The instrument has detected a fault with the barometric sen-
sor or its calibration. This will have an effect on the accu-
racy of the readings as they are pressure compensated. The
instrument will need to be returned to the manufacturer for
service.
Note
*Certain configuration problems can be corrected remotely. Using the G100 Range
- Download Software, it is possible to export the current configuration and e-mail it to the
manufacturer’s Technical Support or Service Department. Depending on the type of error
it may be possible to correct the configuration file and import it back into the instrument.
Warning
Description
Summary of Contents for G100 Series
Page 1: ...Thermo Fisher Scientific G100 Range CO2 Analyser Operating Manual 50121623 12 8 09 ...
Page 6: ...Manual Guidelines Notes 1 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 34: ...Calibration Calibration Record 7 6 G100 Range CO2 Analyser Thermo Scientific ...
Page 40: ...Service 9 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 42: ...Warranty Policy 10 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 46: ...Event Log 12 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 48: ...Sample Certificate of Calibration 13 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 51: ...Thermo Scientific G100 Range CO2 Analyser 15 1 15 EC Declaration of Conformity ...
Page 52: ...EC Declaration of Conformity 15 2 G100 Range CO2 Analyser Thermo Scientific ...
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ......