Chapter 15
Maintenance and Troubleshooting
Maintaning Gas Sampling Valve
304
Maintaning Gas Sampling Valve
The Gas Sampling Valve is normally be serviced by authorized technical
personnel.
User could be required to install/replace the sampling loop. To perform this
operation, please refer to the following operating sequence.
OPERATING SEQUENCE
How to Install/Replace the Loop of an Automatic Valve
WARNING! Before starting the FOCUS GC MUST BE SWITCHED OFF.
Operate as follows:
Remove the FOCUS GC Top Cover
The top cover is fixed on the main unit with two front screws and two clips.
1. Open the GC oven door.
2. Unscrew the two front screws, as shown in Figure 15-22, which remain on the
main unit.
3. Remove manually the top cover pulling it up from the front, the two clips will
be automatically unlocked.
Figure 15-22. Top Aluminium Cover Removal
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...