Chapter 15
Maintaining a SSL Injector
Maintenance and Troubleshooting
265
Removing the Septum
4. Unscrew the septum cap.
5. Remove the septum holder with septum, then the septum support. See A of
Figure 15-1.
6. Remove the septum from the septum holder (use non-metallic tools). See B of
Figure 15-1.
7. Insert a new septum into the septum holder paying attention that the center
guide of the septum must be turned upward. See B of Figure 15-1.
CAUTION
Use tweezers to avoid touching the septum with your fingers.
Figure 15-1. Replace the Septum
Replacing the Septum
8. Clean the septum support from possible fragments left by the septum and
reinsert it into the injector.
B
A
C