Chapter 13
SOP number: P0473/04/E - 02 May 2007
233
Analytical Conditions Required for TCD-S/SL
(*) In case of TCD with the polyimide coated filaments, set the filament temperature limit to 320 °C.
Table 13-2. TCD-S/SL Analytical Conditions
Parameters Setting
Gases
Carrier Gas: Helium = 30 kPa Constant Pressure
Make-up: Helium = 27.5 ml/min
Reference Gas: Helium = 30 ml/min
Oven Program
Initial Temperature = 50 °C
Initial Time = 1 minute
Ramp 1 = 20 °C/minute
Final Temperature = 190 °C
Final Time = 1 minute
Injector
Operating Mode = Splitless
Temperature = 200 °C
Splitless Time = 0.8 minutes
Split Flow = 60 ml/min
Constant Septum Purge = Yes
Detector
Block Temperature = 200 °C
Transfer Temperature = 190 °C
Constant Filament Temperature = No
Filament Voltage = 10V
Filament Temperature limit = 350 °C (*)
Gain = x 10
Negative Polarity = No
Injected Volume
1
μ
l + needle of Test Mixture
Analog Signal
Output
Chrom-Card Acquisition Frequency = Medium
Digital Signal
Output
Chrom-Card, ChromQuest, Xcalibur Acquisition Frequency = 10 Hz
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...