Chapter 9
Thermal Conductivity Detector (TCD)
Selecting TCD Operating Parameters
178
When using low thermal conductivity gases, the temperatures reached by the
filaments are very high. Table 9-2 contains the experimental filament temperature
values corresponding to the applied voltages when using argon.
For samples with a wide range of concentration percentage (1–100%), use the
following values:
•
detector temperature: 10 °C higher than the maximum temperature reached by
the column oven during the analysis, but not higher than 300 ºC
•
filament voltage: 5 V
Suggested Operating Values
Table 9-3 contains the suggested operating values for the detector as a function of
the concentration range and detector temperature.
Table 9-2. Filament Temperature Values for Argon
Detector Temperature 100 ºC
Values
Voltage (V)
5
6
7
8
9
10
Filament Temperature (ºC)
235
275
315
355
395
435
Table 9-3. Selectable TCD Parameters
Concentration
Range
Detector
Temperature
Filament
Temperature
Filament
Voltage
Mode
ppm—5%
0.5–100%
100 ºC
100 ºC
180 ºC
—
—
5 V
CT
CV
ppm—5%
0.5–100%
180 ºC
180 ºC
270 ºC
—
—
6 V
CT
CV
ppm—5%
0.5–100%
240 ºC
240 ºC
330 ºC
—
—
6 V
CT
CV
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...