Chapter 6
Purged Packed Column Injector (PPKD)
PPKD Injection Techniques
116
Septa
You should use high temperature septa with a longer life expectancy, good
resistance to deformation, and a low bleed level, even at high temperatures. Use
high temperature septa for both manual and automatic injections.
Liners
Two different glass liners can be used for wide-bore capillary columns:
•
2 mm ID
•
4 mm ID
PPKD Injection Techniques
The inlet temperature should be sufficiently high to guarantee the sample
completely vaporizes while avoiding the possible sample component
decomposition.
PPKD Injector Menu
The
INLET
menu includes the operating parameters for the purged packed
injector. The parameters you can edit depend on the operating mode chosen:
Wide
bore
,
Packed
,
Wide bore w/surge
or
Packed w/surge
.
In Main Menu move the cursor until
Inlet
is selected. Press
ENTER
to open
INLET
menu
The
Mode
menu items displays the current operating mode. Use
INC
and
DEC
keys
to modify the selection.
Use this list box to select one of the following inlet operating mode:
INLET (PPKD)
Mode Wide bore
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...