21
Appendix 3. ISO8655 calibration specifications
Thermo Scientific F1-ClipTip specifications according to ISO8655 standard.
Fixed Volume
Inaccuracy
Imprecision
Compatible
µl
µl
%
SD µl
CV% ClipTip
10
±0.120
±1.20
0.080
0.80
12.5 Ext
20
±0.20
±1.00
0.10
0.50
20
25
±0.50 ±2.00
0.20
0.80
50
50
±0.50 ±1.00
0.20
0.40
50
100
±0.80 ±0.80
0.30
0.30
200
200
±1.60 ±0.80
0.60
0.30
200
250
±4.00 ±1.60
1.50
0.60
300
500
±4.00 ±0.80
1.50
0.30
1000
1000
±8.00 ±0.80
3.00
0.30
1000
Range Channel Volume
Inaccuracy
Imprecision Compatible Color
µl
µl
%
SD µl CV% ClipTip
code
1–10 µl
8, 12
10
±0.24
±2.4
0.16
1.6
12.5 Ext
pink
1 ±0.24 ±24 0.16 16
5–50 µl
8, 12
50
±1.0
±2.0
0.4
0.8
50
violet
5 ±1.0 ±20 0.4 8.0
10–100 µl 8, 12
100
±0.80
±0.8
0.30
0.3
200
yellow
10 ±0.80 ±8.0 0.30 3.0
30–300 µl 8, 12
300
±8.0
±2.7
3.0
1.0
300
orange
30 ±8.0 ±26.7 3.0 10.0
Thermo Scientific F1-ClipTip Variable Volume Single Channel
Thermo Scientific F1-ClipTip Fixed Volume Single Channel
Thermo Scientific F1-ClipTip Variable Volume Multichannel
Range
Volume Inaccuracy
Imprecision
Compatible Color
µl
µl
%
SD µl
CV%
ClipTip
code
0.1–2 µl
2
±0.080
±4.0
0.040
2.0
12.5 Ext
pink
0.2 ±0.080 ±40.0 0.040 20.0
0.1 ±0.080 ±80.0 0.040 40.0
1–10 µl
10
±0.120
±1.2
0.080
0.8
12.5 Ext
pink
1
±0.120 ±12 0.080 8.0
2–20 µl
20
±0.20
±1.0
0.10
0.5
20
pink
2 ±0.20 ±10 0.10 5.0
5–50 µl
50
±0.50
±1.0
0.20
0.4
50
violet
5 ±0.50 ±10 0.20 4.0
10–100 µl
100
±0.80
±0.8
0.30
0.3
200
yellow
10 ±0.80 ±8.0 0.30 3.0
20–200 µl
200
±1.60
±0.8
0.60
0.3
200
yellow
20 ±1.60 ±8.0 0.60 3.0
30–300 µl
300
±4.0
±1.3
1.5
0.5
300
orange
30 ±4.0 ±13 1.5 5.0
100–1000 µl 1000
±8.0
±0.8
3.0
0.3
1000
blue
100 ±8.0 ±8.0 3.0 3.0