Smart Accessories
Thermo Scientific
Evolution Pro Spectrophotometer User Guide
5
Only the Smart 7-cell rotary changer is supported in the reference position. With this 7 x 7
configuration each cell in the sample position is measured against a unique reference cuvette
in the reference position. This configuration is useful when the different
concentrations/dilutions of matrix, such as buffer solution, are present in a single batch of
samples.
Installing a Smart Accessory
1.
Open the sample compartment and remove any cell holder or accessory.
2.
Insert the accessory.
a.
Grasp the accessory by the handles and lower it into the sample compartment,
aligning the connector under the baseplate with the connector in the sample
compartment floor.
b.
Line up the two holes at the back of the accessory with pins A and B in the sample
compartment.
Alignment hole B
Alignment hole A
Pin A
Pin B
D-connector
(female)
c.
Press down on the front of the accessory to secure the connection.
The software displays a prompt to initialize the accessory.