Installing the EASY-Spray Series Ion Source
Attaching the Ion Source
Thermo Scientific
EASY-Spray Series Ion Source User Guide
9
Powering the Ion Source for Other nanoLC/MS Systems
Follow this procedure to connect the power supply that is provided with the ion source.
To connect the controller power supply to the ion source
1. Turn the ion source’s temperature knob to the OFF position.
2. If not already connected, connect the detachable input plug and power adapter cable to
the power supply (
3. Connect the power adapter cable to the P-BUS socket located on the bottom of the ion
source as follows:
a. Align the red dot on the plug with the red dot on the right side of the socket
b. Push the plug into the socket.
4. Plug the power supply into an electrical wall outlet.
Attaching the Ion Source
Follow these procedures to install the EASY-Spray Series ion source.
•
To install the ion source onto the MS
•
To install the optional Tee holder under the column holder
CAUTION
To avoid an electric shock, always use the detachable input plug that is
appropriate for your country or region and comes with the power supply.
CAUTION
After completing the power supply connections, route the output cord so that
it is not a trip hazard.
Tip
The EASY-Spray NG ion source (ES082) ships with the Tee holder attached to the
column holder. If you have the EASY-Spray ion source (ES081), you can order the
optional Tee holder to support the venting Tee; see