Installing the EASY-Spray Series Ion Source
Attaching the Ion Source
10
EASY-Spray Series Ion Source User Guide
Thermo Scientific
To install the ion source onto the MS
CAUTION
Before you touch an ion source that was recently in use or touch its
components, allow the system to cool for a minimum of 20 minutes.
CAUTION
• Make sure that you do not accidentally lift the release lever located at the top of the
API source interface, which could interfere with or prevent installation of the ion
source.
• To avoid contaminating the ion transfer tube, do not touch its exposed entrance.
Note
You might need to use a small slotted screwdriver to loosen the screws on the
ion sweep cone. The spray cone is directly behind the ion sweep cone.
Align the source mounting holes with the MS pins and carefully
attach the source. Then, lock the source levers.
Unlock both locking levers
(EASY-Spray NG shown).
2
3
If the ion sweep cone is installed, let it cool to room temperature,
and then remove it by grasping its outer ridges and pulling it off.
1
High voltage (HV)
connector
Spring-loaded bolt delivers
HV to the column