Installing AI 1310/AS 1310 Autosampler
Installing the Sampler Support on the Flash IRMS Elemental Analyzer
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
73
To install the direct injection device
1. If present, disconnect the MAS Plus autosampler from the reactor, and place the autosampler nut
on the stainless steel plate. If the MAS Plus autosampler is not present, remove the reactor fitting
unscrewing the relevant fixing nut.
2. Disconnect the gas connection.
3. Install the direct injection device over the reactor. See
Figure 70.
Direct Injection Device Installation
a.
Mount the septum by using the septum holder provided in the standard outfit.
b. Connect the gas line to the direct injection device.
Installing the Sampler Support on the Flash IRMS Elemental Analyzer
The AI 1310/AS 1310 sampling system is installed on the Flash IRMS Elemental Analyzer by using the
appropriate support provided. See
.
Note
When installing the injection device on the pyrolysis side the insert in the reactor must
be replaced by the stainless steel insert which is provided in the water injection kit.
For detailed information please refer to the corresponding manual.
Septum Holder
Septum
Ferrule
6MB Nut
Combustion Reactor
Tube 2 mm OD
O-ring
Nut
Injector Body