 7.  Slide the whole sample area cover back onto the instrument:

Make sure the white plastic feet under the cover are aligned with the wide opening of slots around the 
sample area. Gently set the cover on the baseplate and slide it all the way to the back wall of the 
instrument. Check that the cover sits properly without any gap around the sample area.

 

 8.  Using a flat-blade screwdriver to lock the two cover latches on the inside of front ledge of the sample 

compartment: turn 90-degrees until they are in an upright position.

 

   

11

DXR3xi Class 3B Conversion Guide

Converting DXR3xi from Class 3B to Class 1 configuration

Summary of Contents for DXR3xi

Page 1: ...exposure to hazardous laser radiation l Set up a laser control area in the location where a Class 3B system is being used l When using or working in the vicinity of any Class 3B laser product each per...

Page 2: ...eated Note For the complete laser safety information refer to the DXR Raman Instruments Site and Safety Guide A PDF copy is included in the electronic documentation set and in the online software Help...

Page 3: ...ocks on the system l The connectors and the laser safety circuit controlling access to the Laser Control Area must be designed and built by personnel who have been trained in laser safety and in the o...

Page 4: ...software 2 Remove the BNC interlock jumpers from the Interlock connectors on the instrument panel Properly store the jumpers in a labeled bag They will be used when you want to convert the instrument...

Page 5: ...the inside of front ledge of the sample compartment turn the cover latch 90 degrees until it is in a horizontal position 6 Disconnect the align calibrate tool and set it aside 5 DXR3xi Class 3B Conver...

Page 6: ...away from instrument 8 Take the laser interlock defeating dongle 085 0402xx CBL CLASS 3B OPERATION from the Class 3B Conversion Kit and install it to the interlock connector at the left side of sample...

Page 7: ...is Class 3B system must wear laser safety eyewear that exceeds the required optical safety density for the laser wavelength in use Two laser safety goggles are provided in the DXR Class 3B Conversion...

Page 8: ...on section in the DXRxi Raman Instrument User Guide for installation and align calibration instructions WARNING Avoid eye injury Never put anything except samples against the tip of the fiber optic pr...

Page 9: ...he ports on the fiber optic launcher Do not twist the probe into a tight radius which can damage the fibers b Place the cap onto the end of the probe and the caps onto the ports on the fiber optic lau...

Page 10: ...he external interlock BNC cables 5 Disconnect the align calibrate tool and set it aside 6 Make sure the two cover latches on the inside of front ledge of the sample compartment are in a horizontal pos...

Page 11: ...seplate and slide it all the way to the back wall of the instrument Check that the cover sits properly without any gap around the sample area 8 Using a flat blade screwdriver to lock the two cover lat...

Page 12: ...to Class 1 configuration 10 Follow normal Align Calibration procedure and make sure all tests pass 11 Refer to the Class 3B Conversion Guide in the electronic documentation set for future reference 1...