Maintaining the DSQ II
Replacing the Electron Multiplier
Thermo Scientific
DSQ II Hardware Manual
61
Figure 35.
Ion Detector Components
a.
Remove the two screws that hold the shield in place.
b. Remove the two screws that hold the base
in place.
c.
Disconnect the electron multiplier wire.
d. Remove the old electron multiplier by gently pressing on the cathode springs while
pulling up on the electron multiplier.
7. Lock the new electron multiplier in place with springs and clip.
8. Use the two base screws to reattach the ion detector base to the manifold wall. Ensure the
conversion dynode connector wire is seated properly in the dynode and feedthrough.
9. Reassemble the ion detector.
a.
Attach the shield with the two screws that hold the shield in place.
b. Attach the new electron multiplier cathode with the spring clip. Insert the end nearest
the manifold mount first, then push the other end into the clip.
c.
Reconnect the electron multiplier connector.
10. Reattach the front and top covers.
11. Restart the DSQ II.
1
Electron Multiplier Cathode
5
Electron Multiplier Wire
2
Conversion Dynode Wire
6
Base
3
Two Screws for Shield
7
Two Screws for the Base
4
Shield and Conversion Dynode
8
Electron Multiplier Anode
2
6
5
3
1
4
7
8
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...