5
Ion Source and Inlet Valve
Ion Source
170
DSQ II Hardware Manual
Thermo Scientific
Ion Source
During operation, the ion source holds an ion volume in its center. It is aligned by the ion
volume key thumbscrew and secured by the ball plunger. Samples are introduced from a GC
through the transfer line into the sample inlet aperture located on the side of the ion source.
Calibration compound and chemical ionization reagent gas also enter the ion volume through
the transfer line.
The ion source board supplies voltages to the filament, electron lens, and cartridge heaters, as
it passes signals from the temperature sensor. Three base studs hold the ion source board to
the ion source. These studs slip into the magnet yoke and position the ion source between two
permanent magnets.
Three cartridge heaters and a temperature sensor are mounted on the ion source board.
Cartridge heaters pass through and heat the ion source, as well as heat, align, and support the
three ion source lenses. The temperature sensor monitors the temperature of the ion source.
The heater ring slips over the cartridge heaters and transfers heat to the lenses.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...