2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
94
DSQ II Hardware Manual
Thermo Scientific
2. Remove and replace the inlet valve seal.
Figure 53.
Removing the Inlet Valve Seal
a.
Insert the inlet seal removal tool into the inlet valve.
b. Press the button on the tool to engage the inlet valve seal.
c.
Pull out the inlet valve seal and discard it.
d. Insert a new inlet valve seal (PN 119265-0003).
3. Reassemble the inlet valve.
a.
Put the ferrule, knob and plug back into the inlet valve opening.
b. The first time you use the inlet valve, you will need to tighten the inlet valve knob
more than usual to prevent it from leaking.
Replacing the Ion Gauge (Optional Upgrade)
To replace the ion gauge, you will need:
• Gloves, clean, lint- and powder-free
• Ion gauge (PN A0105-06003)
• Wrench, Allen, 3 mm
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
Inlet Seal Removal Tools
CAUTION - INSTRUMENT DAMAGE.
To avoid scratching the sealing surface, you should
only use the supplied inlet seal removal tool to remove the seal.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...