Dionex Integral Systems Operator’s Manual
72
Doc. 065250-06 03/2020
5.4
AE Power and Computer Connections
Power Receptacle
The power in receptacle on the top of the AE (see
connection to the AE main power. Connect the power cord provided in the AE
Ship Kit (P/N 069066) from this connector to a grounded, single-phase power
source.
The power cord is configured with a NEMA L5-20P plug (125 Vac/20 A twist-
lock) (see
) for connection to facility power terminated as a NEMA L5-
20R wall receptacle. Other wall plugs or hard-wired connections may be used,
provided that arrangements are made with Thermo Fisher Scientific before
Figure 5-4. AE Power In Receptacle
Power In
Receptacle