Dionex Integral Systems Operator’s Manual
26
Doc. 065250-06 03/2020
The
Mixer Stir Bar Auto-Recovery
option on the
Pumps/Motors
page
uses Analog Input 4 to monitor the mixer stir bar. If the stir bar is in an
out-of-control state, the mixer motor is stopped briefly and then restarted
at a speed 5% slower than before the auto-recovery occurred. To disable
auto-recovery, select
No
for this option.
You can use the controls on the SP ePanel in Chromeleon 7 to turn the
mixer on and off and to set the speed.
Heated Dilution Vessel (Option)
The temperature of the heated dilution vessel can be set to between 15 °C
and 40 °C. Specify the temperature in the Chromeleon 7 Instrument
Configuration Manager on the
Temperature
page or in the Chromeleon 7
instrument method.
3.5.3
Dilution Pump
The dilution pump (P/N 068561) (see
) uses a stepper motor to precisely
deliver diluent (typically deionized water)
through the metering valve to the dilution
vessel. The pump can deliver from 0.1 mL
to 250.0 mL.
The pump flow rate can be set to between
0.0 mL/min and 15.0 mL/min (the default is
15.0 mL/min). You can change the default
flow rate in the Chromeleon 7 Instrument
Configuration Manager on the
Pumps/Motors
page.
Figure 3-6. Dilution Pump
Enter commands for control of the pump in the Chromeleon 7 instrument
method or on the SP ePanel in Chromeleon 7.
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...