9.3
Liquid Leaks/Leak Alarm
NOTE When cutting tubing and preparing fittings, avoid
crimping the tubing. Crimped tubing is a common cause
of high backpressure.
NOTE After eliminating the source of a leak, always dry the
leak sensor thoroughly. If the leak sensor is not dry, it
will remain activated and will continue to report a leak
to the audit trail.
•
Leaking fitting
Locate the source of the leak. Tighten or, if necessary, replace the liquid line
connection.
•
Broken or damaged liquid line
Cut the tubing at the break and install a new fitting.
–or–
If cutting the tubing would make it too short, replace it. The new tubing must
be the same type (and have the same internal diameter) as the tubing it
replaces.
–or–
If leaking liquid creates a hazard, stop the leak immediately by turning
off the flow at the source.
Si une fuite de liquide crée un danger, arrêtez immédiatement la fuite
en fermant l'écoulement à la source.
Wenn eine Gefährdung durch austretende Flüssigkeit besteht,
stoppen Sie die Leckage unmittelbar, indem Sie den Fluß an der
Quelle abstellen.
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...