This error occurs if the dilution pump electronics fail to detect when the pump
reaches the home position. This can be caused by high pressure in the flow
path.
To troubleshoot:
1. Check for a blockage in the pump outlet tubing or in the tubing
for the SP1 flow schematic;
2. If no blockage is found, turn off the pump flow.
3. Turn off the SP power briefly by pressing the SP power switch.
4. Turn on the SP power again and restart the pump flow.
5. If the problem persists, contact Technical Support for Dionex products for
assistance.
This is a warning message only. The pump continues to function and no action
is needed.
This error can be caused by high pressure in the flow path.
To troubleshoot:
1. Check for a blockage in the pump outlet tubing or in the tubing
for the SP1 flow schematic;
2. If no blockage is found, turn off the pump flow.
3. Turn off the SP power briefly by pressing the SP power switch.
4. Turn on the SP power again and restart the pump flow.
Dilution pump home sensor not detected
Dilution pump lost count
Dilution pump stopped or lost more than 20 steps
Summary of Contents for Dionex
Page 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Page 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Page 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Page 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Page 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Page 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Page 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Page 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Page 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Page 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Page 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...